⛽ Hydrocarbon Oil Index in Water — Petroleum Hydrocarbons (GC-FID)
In short
Determination of the hydrocarbon oil index in surface waters, waste waters and water from sewage treatment plants at concentrations above 0. The test is performed according to PN-EN ISO 9377-2:2003; Field of application: surface waters, waste waters, water from treatment plants; index above 0.1 mg/l; not volatile mineral oils. The procedure comprises 7 steps; it is used for: Surface waters — hydrocarbon oil index above 0.1 mg/l, Waste waters and water from sewage treatment plants — with a limit for animal and vegetable fats (up to 150 mg/l) and a risk of low recovery with large amounts of surfactants, Groundwaters — with field preservation, unless a high content of humic substances rules out acidification.
At a glance
- Standard: PN-EN ISO 9377-2:2003
- Category: Physicochemistry
- Procedure steps: 7
- Field of application: surface waters, waste waters, water from treatment plants; index above 0.1 mg/l; not volatile mineral oils
- Retention window: from n-decane (C₁₀) to n-tetracontane (C₄₀)
- Extractant: hydrocarbon solvent or technical mixture with a boiling point of 36–69 °C; 50 ml per sample of approx. 900 ml
Overview
STATUS (as of 28.10.2026). PN-EN ISO 9377-2:2003 — Polish version, published 17.10.2003, implements EN ISO 9377-2:2000 and ISO 9377-2:2000 [IDT]; it replaced PN-EN ISO 9377-2:2002 (English version). The standard is current — no withdrawal note.
SCOPE (clause 1). The method is used to determine the hydrocarbon oil index in water by gas chromatography; it is suitable for surface water, waste water and water from sewage treatment plants and allows the index to be determined at concentrations above 0.1 mg/l. It is not suitable for the quantitative determination of volatile mineral oils, but the composition of the contamination can be inferred from the appearance of the chromatogram. Mineral oil content in soil and sediments — according to ISO/TR 11046. The concentration of animal and vegetable fats in the sample should not exceed 150 mg/l, because at higher levels the capacity of the clean-up column may be insufficient; in highly polluted waste waters, especially those with large amounts of surfactants, the recovery may be low.
DEFINITION (3.1). The hydrocarbon oil index determined by GC-FID is the sum of the concentrations of compounds extractable with a hydrocarbon solvent with a boiling point of 36–69 °C, not adsorbed on Florisil and eluting with retention times between n-decane (C₁₀H₂₂) and n-tetracontane (C₄₀H₈₂). These are long-chain or branched aliphatic, alicyclic, aromatic and alkyl-substituted aromatic hydrocarbons. The result is therefore defined by the course of the method itself.
INTERFERENCES (clause 4). Compounds of low polarity (e.g. halogenated hydrocarbons) and high concentrations of polar substances interfere; surfactants interfere with the extraction.
CALIBRATION AND STANDARDS (clause 6). The external standard is a mixture of equal amounts of two mineral oils without additives: type A — with distinct peaks, e.g. diesel fuel, and type B — with a higher boiling range and an unresolved signal, e.g. a lubricating oil with a boiling range of 325–460 °C; together approx. 10 mg/ml. At least five calibration solutions, e.g. 0; 0.2; 0.4; 0.6; 0.8 and 1.0 mg/ml, stable for up to six months at 4–8 °C. The extraction solvent contains n-decane and n-tetracontane as markers of the limits of the retention window. A mixture of n-alkanes (C₂₀, C₄₀ and at least three others) is used to check the chromatographic system, and a stearyl stearate solution to check the Florisil.
CONTROL REQUIREMENTS (clause 9). A blank in every series; recovery from water spiked with the control standard, preferably in every series, within 80–110 %; suitability of every new batch of Florisil — stearyl stearate must be retained and the recovery of mineral oil through the column must be at least 80 %; the n-alkane peaks resolved to the baseline, and the response of n-tetracontane relative to n-eicosane at least 0.8 (otherwise the injector discriminates against heavy hydrocarbons).
WHAT WE DO NOT GIVE HERE. The chromatographic conditions (an example in Annex B), details of the calibration, the calculation formulas and the precision data (9.7.2 onwards) — we only know their titles.
Method principle
The water sample (acidified to pH 2) is extracted in the bottle in which it was collected with a hydrocarbon solvent with a boiling point of 36–69 °C containing n-decane and n-tetracontane, stirring for 30 minutes with a magnetic stirrer; magnesium sulfate prevents emulsion. Polar substances are removed from the extract on a column of 2 g of Florisil under 2 g of anhydrous sodium sulfate, the extract is concentrated to 1 ml and analysed by capillary gas chromatography on a non-polar column with a flame ionisation detector.
The total peak area between n-decane and n-tetracontane is measured and the hydrocarbon oil index is calculated against an external standard of two specified mineral oils. The mass of the sample is determined by weighing the bottle before extraction and after emptying it.
Applications
- Surface waters — hydrocarbon oil index above 0.1 mg/l
- Waste waters and water from sewage treatment plants — with a limit for animal and vegetable fats (up to 150 mg/l) and a risk of low recovery with large amounts of surfactants
- Groundwaters — with field preservation, unless a high content of humic substances rules out acidification
- Qualitative assessment of the composition of mineral oil contamination from the appearance of the chromatogram
Key parameters
| Parameter | Value |
|---|---|
| Field of application | surface waters, waste waters, water from treatment plants; index above 0.1 mg/l; not volatile mineral oils |
| Retention window | from n-decane (C₁₀) to n-tetracontane (C₄₀) |
| Extractant | hydrocarbon solvent or technical mixture with a boiling point of 36–69 °C; 50 ml per sample of approx. 900 ml |
| Clean-up | 2 g of Florisil (150–250 µm, 16 h at 140 °C) under 2 g of anhydrous sodium sulfate |
| GC column | non-polar (100 % dimethylpolysiloxane, 95/5 dimethyl/diphenyl polysiloxane or a modified siloxane polymer), 5–30 m, diameter 0.25–0.53 mm, film 0.25–1.2 µm |
| Sample storage | approx. 4 °C, extraction as soon as possible, at the latest within 4 days |
| Recovery (control) | 80–110 % |
| Fats in the sample | not more than 150 mg/l |
Standard
- Standard number
- PN-EN ISO 9377-2:2003
- Title (PL)
- Jakość wody — Oznaczanie indeksu oleju mineralnego — Część 2: Metoda z zastosowaniem ekstrakcji rozpuszczalnikiem i chromatografii gazowej
- Title (EN)
- Water quality — Determination of hydrocarbon oil index — Part 2: Method using solvent extraction and gas chromatography
Step-by-step procedure
-
Sampling and storage
According to ISO 5667-3. Fill a glass bottle to approx. 90 %, close it and weigh it (m1). Store at approx. 4 °C and extract as soon as possible, at the latest within 4 days. If necessary, preserve in the field with acid to pH 2 — recommended for surface waters and groundwaters, but not with a high content of humic substances (yellow-brown colour). If an emulsion or fats above 150 mg/l are expected, additionally take a smaller sample in a 250 ml bottle (clause 8).
-
Blank and recovery
A blank with every series, with all reagents and glassware. Recovery — preferably in every series: 900 ml of water with 1.0 ml of control standard, the entire procedure; recovery 80–110 % (9.1–9.2).
-
Extraction
If necessary, cool the sample to approx. 10 °C, acidify to pH 2 (if not done in the field), add approx. 80 g of magnesium sulfate per 900 ml (not needed if the sample does not form an emulsion), 50 ml of extraction solvent with the standards and a stirring bar; stir vigorously for 30 min. Collect the solvent layer with a microseparator, avoiding transfer of water; break strong emulsions by centrifugation. Extraction by shaking is permitted after its efficiency has been checked (9.3).
-
Clean-up on Florisil
Pass the extract through a column of 2 g of Florisil under 2 g of sodium sulfate, then an additional 10 ml of solvent; rinse the column with approx. 10 ml. Shaking the whole extract with 2 g of Florisil for 30 min is permitted if the results are equivalent (9.4).
-
Concentration
Concentrate to approx. 6 ml in a Kuderna-Danish apparatus or an evaporator, then with a gentle stream of nitrogen to slightly below 1 ml and make up to 1.0 ml (or determine the volume by weighing). With a high expected index or injection of large volumes, concentration may be omitted. Dry the empty bottle for 5 min and weigh it (m2) to the nearest 1 g (9.5).
-
Checking the Florisil
Periodically and with every new batch: stearyl stearate must be retained and the recovery of mineral oil through the column must be at least 80 %; otherwise wash and reactivate the Florisil or use another batch (9.6).
-
Gas chromatography
Choose the conditions so that the peaks of the n-alkane mixture are resolved to the baseline and the response of n-tetracontane relative to n-eicosane is at least 0.8 — otherwise improve or replace the injection system (9.7.1). Calibration, integration of the area between C₁₀ and C₄₀ and calculation of the index — according to the further clauses of the standard.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Gas chromatograph with a flame ionisation detector | with a non-discriminating injection system; a data system for integrating the chromatogram area with compensation for column bleed (7.2, 7.4) | — |
| Capillary column and pre-column | fused silica, non-polar, 5–30 m × 0.25–0.53 mm; a pre-column is recommended, e.g. 2 m × 0.53 mm of deactivated fused silica (7.3) | — |
| Glass sampling bottles | 250 ml and 1000 ml, with a ground glass stopper or a screw cap with a PTFE liner; extraction directly in the bottle (7.5) | — |
| Magnetic stirrer and microseparator | for extraction in the bottle and separation of the solvent layer; centrifuge and 100 ml tubes for breaking emulsions (7.6–7.8, 7.11) | — |
| Clean-up columns | glass with a sintered disc of porosity 2 (7.9) | — |
| Concentration apparatus | Kuderna-Danish apparatus with a 250 ml flask or a rotary evaporator with controlled vacuum; then a gentle stream of nitrogen (7.10, 9.5) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Extraction solvent | — | a single hydrocarbon solvent or a technical mixture with a boiling point of 36–69 °C; a change of solvent requires repeatability tests (6.2); the working solution contains n-decane and n-tetracontane (6.11) |
| Florisil | — | particle size 150–250 µm (60–100 mesh), heated for 16 h at 140 °C, stored in a desiccator (6.7) |
| Anhydrous sodium sulfate and magnesium sulfate heptahydrate | — | sodium sulfate above the Florisil in the column; magnesium sulfate approx. 80 g per 900 ml of sample against emulsion (6.3–6.4, 9.3) |
| Mineral acid | — | e.g. hydrochloric acid 12 mol/l — for acidifying the sample to pH 2 (6.5) |
| Mixture of mineral oils of type A and B | — | equal amounts of type A oil (e.g. diesel fuel without additives) and type B oil (e.g. lubricating oil without additives, boiling range 325–460 °C), approx. 10 mg/ml — calibration and control standard (in acetone) (6.6, 6.8) |
| n-alkane mixture and stearyl stearate | — | n-alkanes with an even number of carbon atoms (C₂₀, C₄₀ and at least three others) for checking the GC system; stearyl stearate for checking the Florisil (6.9, 6.12) |
Health and safety (OHS)
- The extraction solvent (boiling point 36–69 °C) is extremely flammable — work in a fume hood, away from ignition sources
- Concentrated hydrochloric acid (12 mol/l) is corrosive — follow the safety data sheet
- Hydrogen for the flame ionisation detector — check the tightness of the chromatograph gas installation
- The test requires suitably qualified staff (clause 5)
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN ISO 9377-2:2003 — “Water quality — Determination of hydrocarbon oil index — Part 2: Method using solvent extraction and gas chromatography”.
How does this method work?
The water sample (acidified to pH 2) is extracted in the bottle in which it was collected with a hydrocarbon solvent with a boiling point of 36–69 °C containing n-decane and n-tetracontane, stirring for 30 minutes with a magnetic stirrer; magnesium sulfate prevents emulsion. Polar substances are removed from the extract on a column of 2 g of Florisil under 2 g of anhydrous sodium sulfate, the extract is concentrated to 1 ml and analysed by capillary gas chromatography on a non-polar column with a flame ionisation detector.
What is the measuring range and accuracy?
Field of application: surface waters, waste waters, water from treatment plants; index above 0.1 mg/l; not volatile mineral oils; Retention window: from n-decane (C₁₀) to n-tetracontane (C₄₀); Extractant: hydrocarbon solvent or technical mixture with a boiling point of 36–69 °C; 50 ml per sample of approx. 900 ml; Clean-up: 2 g of Florisil (150–250 µm, 16 h at 140 °C) under 2 g of anhydrous sodium sulfate.
How long does the test take?
The procedure comprises 7 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Gas chromatograph with a flame ionisation detector, Capillary column and pre-column, Glass sampling bottles, Magnetic stirrer and microseparator, Clean-up columns, Concentration apparatus.
Where is this test used?
Surface waters — hydrocarbon oil index above 0.1 mg/l; Waste waters and water from sewage treatment plants — with a limit for animal and vegetable fats (up to 150 mg/l) and a risk of low recovery with large amounts of surfactants; Groundwaters — with field preservation, unless a high content of humic substances rules out acidification; Qualitative assessment of the composition of mineral oil contamination from the appearance of the chromatogram.
What safety precautions apply?
The extraction solvent (boiling point 36–69 °C) is extremely flammable — work in a fume hood, away from ignition sources; Concentrated hydrochloric acid (12 mol/l) is corrosive — follow the safety data sheet; Hydrogen for the flame ionisation detector — check the tightness of the chromatograph gas installation; The test requires suitably qualified staff (clause 5).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 9377-2.