🏭 Volatile organic compounds in emissions
In short
Determination of the mass concentration of individual gaseous organic compounds in waste gases, for example from solvent using processes: sampling onto activated carbon tubes, solvent desorption and analysis by capillary gas chromatography; the withdrawn standard covers only solvent desorption, and thermal desorption was added only by the Technical Specification CEN/TS 13649:2014. The test is performed according to PN-EN 13649:2005; Concentration range: about 0.5–2000 mg/m³ for individual components (EN 13649:2001). The procedure comprises 7 steps; it is used for: Mass concentration of individual gaseous organic compounds in waste gases, e.g. from solvent using processes (EN 13649:2001, clause 1), Range from about 0.5 mg/m³ to 2000 mg/m³ (EN 13649:2001, clause 1), Installations using organic solvents — VOC emission standards (Poland: Dz.U. 2020 item 1860, Chapter 6).
At a glance
- Standard: PN-EN 13649:2005
- Category: Environmental
- Procedure steps: 7
- STATUS OF THE DOCUMENT: PN-EN 13649:2005 WITHDRAWN 08-08-2018; no successor in the PN collection; EN 13649:2001 superseded by CEN/TS 13649:2014
- European standard implemented: EN 13649:2001 [IDT] — publication date of the Polish edition 28-04-2005
- Concentration range: about 0.5–2000 mg/m³ for individual components (EN 13649:2001)
Overview
STATUS (as of 06.11.2026). PN-EN 13649:2005 — Polish version, published on 28.04.2005, which implemented EN 13649:2001 [IDT] — has been WITHDRAWN since 08.08.2018; the PKN card does not name a replacing standard and there is no successor in the Polish Standards collection. At European level EN 13649:2001 was superseded by the Technical Specification CEN/TS 13649:2014, which has not been adopted into the PN collection. Despite the withdrawal, the standard has 386 entries in Polish scopes of accreditation at 81 laboratories (copy of the scope database, 06.11.2026) — citing a withdrawn document requires a deliberate decision agreed with the customer.
SCOPE (EN 13649:2001, clause 1). The standard specifies the sampling onto activated carbon, the preparation and the analysis of samples of volatile organic components, e.g. from solvent using processes (a note refers to Directive 1999/13/EC); it can be used as a reference method. The result is the mass concentration (mg/m³) of individual gaseous organic components, in the range from about 0.5 mg/m³ to 2000 mg/m³. For total organic carbon from solvent using processes the standard refers to EN 13526.
PRINCIPLE (clause 4). Three steps: flue gas sampling, treatment of the sampled material and analysis by gas chromatography. Organic components from a measured volume of gas are adsorbed onto activated carbon; interfering particulate matter should be removed and water condensation prevented. The collected material is desorbed with an extraction solvent, an aliquot is injected into a gas chromatograph with a capillary column and an appropriate detector (e.g. flame ionization or mass selective), and evaluated using the calibration function. Desorption efficiency is determined by adding a known amount of reference material to a blank tube; according to a note in the standard it should be better than 80 %.
SAMPLING (clauses 5–6). A system of chemically and physically inert material (a note lists stainless steel, glass, PTFE and polypropylene fluoride as well proven), with a sampling line as short as possible, a filter directly after the probe and protection against condensation. The standard shows examples of sampling without dilution, with dynamic dilution (dilution constant to within 5 %) and with static dilution (bag of inert material in a rigid container; pressure control device not above 250 Pa). The sorbent tube has a main layer of at least 100 mg of activated carbon (e.g. from coconut shell) and a security layer to detect breakthrough; open or used tubes are not reused. Dilution gas — nitrogen or dry air containing less than 0.1 % of the emission limit value of the components investigated. The gas temperature during adsorption should not exceed 40 °C; when the gas is so moist that water condenses, or the concentration risks exceeding the tube capacity, dilution is used. The sample shall represent the process emission and the distribution in the sampling plane; the sampling time is chosen e.g. by the required averaging period and the detection limit.
REAGENTS (5.2). Extraction solvent of chromatographic quality, free from components co-eluting with those of interest; carbon disulfide (CS₂) is suitable for most components from solvent using processes, other solvents for polar substances (Annex B). The chromatograph is calibrated with reference materials corresponding to the process components, prepared in a solution of the extraction solvent; standards should be prepared fresh regularly.
CEN/TS 13649:2014 — WHAT CHANGED. The Technical Specification superseded EN 13649:2001: status changed from standard to Technical Specification, clarified scope, a decision tree for the sampling procedure, sampling strategy aligned with EN 15259, thermal desorption added, and validation information. The solvent-extraction tube has 100 mg of activated carbon in the main layer and 50 mg in the security layer; thermal desorption tubes (stainless steel, inert-coated steel or glass) can be reused. Pump flow — e.g. up to 0.1 l/min for thermal desorption tubes and up to 0.5 l/min for charcoal tubes; thermal desorption is typically three orders of magnitude more sensitive. Liquid water shall not reach the sorbent, and drying tubes (e.g. sodium sulfate) are not used upstream of the sorbent. According to a note in the specification, the limit values of EU Directives are expressed in mg/m³ on a wet basis for non-combustion processes and on a dry basis for combustion, at the reference conditions of 273 K and 101.3 kPa.
REGULATIONS (Poland). Emission standards for installations using organic solvents are set by Chapter 6 of the Regulation of the Minister of Climate of 24 September 2020 on emission standards (Dz.U. 2020 item 1860, § 30–31, Annexes 9–10); it repealed the 2018 regulation (Dz.U. 2018 item 680).
WHAT WE DO NOT GIVE HERE. The minimum sampling time and flow rate of EN 13649 (6.3 onwards), the desorption procedure, chromatographic conditions, calculations and uncertainty — they are outside the available text of both documents. Temperature programmes of the desorber and the chromatograph or trade names of tubes — the sources do not give them.
Method principle
Waste gas is sampled through a filter and a line of inert material, preventing condensation (with dilution if needed), onto an activated carbon tube with a main and a security layer. The organic components are desorbed with an extraction solvent (usually carbon disulfide), an aliquot is analysed by capillary gas chromatography with a detector, for example flame ionization or mass selective, and the concentration is determined from the calibration function taking account of the desorption efficiency. The Technical Specification CEN/TS 13649:2014, which superseded the standard, also allows tubes with sorbents for thermal desorption.
Applications
- Mass concentration of individual gaseous organic compounds in waste gases, e.g. from solvent using processes (EN 13649:2001, clause 1)
- Range from about 0.5 mg/m³ to 2000 mg/m³ (EN 13649:2001, clause 1)
- Installations using organic solvents — VOC emission standards (Poland: Dz.U. 2020 item 1860, Chapter 6)
- Monitoring of individual VOCs according to the requirements of the Industrial Emissions Directive — according to CEN/TS 13649:2014 (clause 1)
- Not for total organic carbon — for it EN 13526 (formerly) and EN 12619 (according to CEN/TS 13649)
Key parameters
| Parameter | Value |
|---|---|
| STATUS OF THE DOCUMENT | PN-EN 13649:2005 WITHDRAWN 08-08-2018; no successor in the PN collection; EN 13649:2001 superseded by CEN/TS 13649:2014 |
| European standard implemented | EN 13649:2001 [IDT] — publication date of the Polish edition 28-04-2005 |
| Concentration range | about 0.5–2000 mg/m³ for individual components (EN 13649:2001) |
| Sorbent tube | main layer of at least 100 mg activated carbon + security layer (EN 13649:2001); 100 mg + 50 mg (CEN/TS 13649:2014) |
| Gas temperature during adsorption | should not exceed 40 °C (6.2) |
| Desorption efficiency | should be better than 80 % (4.5, note) |
| Flow rate (CEN/TS 13649:2014) | e.g. up to 0.5 l/min for charcoal tubes and up to 0.1 l/min for thermal desorption tubes |
Standard
- Standard number
- PN-EN 13649:2005
- Title (PL)
- Emisja ze źródeł stacjonarnych — Oznaczanie stężenia masowego indywidualnych gazowych związków organicznych — Metoda z zastosowaniem węgla aktywnego i desorpcji rozpuszczalnikiem
- Title (EN)
- Stationary source emissions — Determination of the mass concentration of individual gaseous organic compounds — Activated carbon and solvent desorption method
Step-by-step procedure
-
Measurement plan
Ensure the sample is representative of the process emission and the distribution in the sampling plane; if possible measure under steady operating conditions of the plant (6.1; strategy according to EN 15259 in CEN/TS 13649:2014).
-
Choice of sampling procedure
Sampling directly onto the sorbent when the gas is not very moist and breakthrough is unlikely; static or dynamic dilution when water condenses or the concentration risks exceeding the tube capacity (6.2).
-
Sampling
Sample a measured volume of gas through the filter and line without condensation onto a tube with main and security layers; the gas temperature during adsorption should not exceed 40 °C; choose the sampling time according to the averaging period and the detection limit (4.2, 6.2–6.3).
-
Desorption
Desorb the collected components with the extraction solvent (4.3); according to CEN/TS 13649:2014 alternatively thermal desorption.
-
Desorption efficiency
Add a known amount of reference material to a blank tube and determine the desorption efficiency (should be better than 80 %) (4.5, Annex B).
-
Analysis
Inject an aliquot into the chromatograph with a capillary column and an appropriate detector; evaluate using the calibration function prepared from reference materials (4.4, 5.2.3).
-
Calculation and report
Express the result as the mass concentration of individual components in mg/m³ according to clauses 8–9 of the standard.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Probe with prefilter and sampling line | of inert material (stainless steel, glass, PTFE, polypropylene fluoride), as short as possible, with a filter directly after the probe, without condensation (5.1) | — |
| Activated carbon sorbent tubes | main layer of at least 100 mg and a security layer, glass, free from contamination, single use (5.1.4) | — |
| Pump, adjustment valve, flow meter, gas meter and drying unit | components of the example sampling system (Figures 1–2) | — |
| Dynamic or static dilution system | dilution constant to within 5 %; bag in a rigid container with a pressure control device up to 250 Pa (5.1.2–5.1.3) | — |
| Gas chromatograph with capillary column | with an appropriate detector, e.g. flame ionization or mass selective (4.4, 5.3) | — |
| Thermal desorber | only according to CEN/TS 13649:2014 — for tubes with thermal desorption sorbents | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Carbon disulfide (CS₂) | 75-15-0 | extraction solvent suitable for most components; chromatographic quality, free from co-eluting components (5.2.2) |
| Activated carbon | 7440-44-0 | tube adsorbent, e.g. from coconut shell (5.1.4) |
| Reference materials | — | corresponding to the process components, in a solution of the extraction solvent, prepared fresh regularly (5.2.3) |
| Dilution gas | — | nitrogen or dry air with less than 0.1 % of the emission limit value of the components investigated; e.g. nitrogen 99.999 % (5.1.5) |
Health and safety (OHS)
- Carbon disulfide — harmonised CLP classification: H225, H315, H319, H361fd, H372; work under a fume hood, away from ignition sources
- The standard contains no other safety requirements; the rules for working at measurement sites on the stack are set by the laboratory on the basis of a risk assessment
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 13649:2005 — “Stationary source emissions — Determination of the mass concentration of individual gaseous organic compounds — Activated carbon and solvent desorption method”.
How does this method work?
Waste gas is sampled through a filter and a line of inert material, preventing condensation (with dilution if needed), onto an activated carbon tube with a main and a security layer. The organic components are desorbed with an extraction solvent (usually carbon disulfide), an aliquot is analysed by capillary gas chromatography with a detector, for example flame ionization or mass selective, and the concentration is determined from the calibration function taking account of the desorption efficiency.
What is the measuring range and accuracy?
STATUS OF THE DOCUMENT: PN-EN 13649:2005 WITHDRAWN 08-08-2018; no successor in the PN collection; EN 13649:2001 superseded by CEN/TS 13649:2014; European standard implemented: EN 13649:2001 [IDT] — publication date of the Polish edition 28-04-2005; Concentration range: about 0.5–2000 mg/m³ for individual components (EN 13649:2001); Sorbent tube: main layer of at least 100 mg activated carbon + security layer (EN 13649:2001); 100 mg + 50 mg (CEN/TS 13649:2014).
How long does the test take?
The procedure comprises 7 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Probe with prefilter and sampling line, Activated carbon sorbent tubes, Pump, adjustment valve, flow meter, gas meter and drying unit, Dynamic or static dilution system, Gas chromatograph with capillary column, Thermal desorber.
Where is this test used?
Mass concentration of individual gaseous organic compounds in waste gases, e.g. from solvent using processes (EN 13649:2001, clause 1); Range from about 0.5 mg/m³ to 2000 mg/m³ (EN 13649:2001, clause 1); Installations using organic solvents — VOC emission standards (Poland: Dz.U. 2020 item 1860, Chapter 6); Monitoring of individual VOCs according to the requirements of the Industrial Emissions Directive — according to CEN/TS 13649:2014 (clause 1); Not for total organic carbon — for it EN 13526 (formerly) and EN 12619 (according to CEN/TS 13649).
What safety precautions apply?
Carbon disulfide — harmonised CLP classification: H225, H315, H319, H361fd, H372; work under a fume hood, away from ignition sources; The standard contains no other safety requirements; the rules for working at measurement sites on the stack are set by the laboratory on the basis of a risk assessment.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 13649.