🌫️ PM10 / PM2.5 particulate matter concentration in air
In short
Standard gravimetric method for the mass concentration of suspended particulate matter in ambient air: the particulate matter is collected on a filter through a size-selective inlet over a nominal 24 h and the filter is weighed before and after sampling under fixed conditions. The test is performed according to PN-EN 12341:2024-01; Range of application (24 h): from approximately 1 µg/m³ to 150 µg/m³ (PM10) and 120 µg/m³ (PM2.5) (clause 1). The procedure comprises 9 steps; it is used for: Determination of PM10 or PM2.5 concentration in ambient air — rural areas, urban background, traffic-orientated locations and locations influenced by industrial sources (clause 1), Assessment of compliance with limit values: PM10 50 µg/m³ for a day and 40 µg/m³ for a year, PM2.5 20 µg/m³ for a year (Poland: Dz.U. 2021 item 845), Reference methodology for assessing levels of suspended particulate matter (Poland: Dz.U. 2024 item 870, Annex 7 — 2014 edition; Directive (EU) 2024/2881 — 2023 edition).
At a glance
- Standard: PN-EN 12341:2024-01
- Category: Environmental
- Procedure steps: 9
- Flow rate: 2.3 m³/h — nominal (clause 1)
- Sampling period: nominally 24 h (clause 1, 4.1)
- Range of application (24 h): from approximately 1 µg/m³ to 150 µg/m³ (PM10) and 120 µg/m³ (PM2.5) (clause 1)
Overview
STATUS (as of 06.11.2026). PN-EN 12341:2024-01 — English version, published on 29.01.2024, implements EN 12341:2023 [IDT]; it replaced PN-EN 12341:2014-07 (English version), withdrawn on the same day. The PKN card has no withdrawal note — the standard is current. EN 12341:2023 superseded EN 12341:2014; the technical changes are listed in Annex I of the standard. The Polish regulation on the assessment of substance levels in air still cites the PN-EN 12341:2014-07 edition, while the new EU ambient air quality directive cites EN 12341:2023 (details below).
SCOPE (clause 1). The standard specifies a standard method for determining the PM10 or PM2.5 mass concentration of suspended particulate matter in ambient air by sampling the particulate matter on filters and weighing them. Measurements are performed with samplers with inlets as specified in Annex A, at a nominal flow rate of 2.3 m³/h and a nominal sampling period of 24 h, in rural areas, urban background, traffic-orientated locations and locations influenced by industrial sources. Results are expressed in µg/m³, with the air volume at ambient conditions near the inlet at the time of sampling. The range of application for 24 h measurements is from approximately 1 µg/m³ (the limit of detection expressed as uncertainty) up to 150 µg/m³ for PM10 and 120 µg/m³ for PM2.5; a note allows extension to about 200 µg/m³ with suitable filters. The standard mainly concerns sequential samplers with a filter changer but does not exclude single-filter samplers.
DEFINITION (3.1.16). PMx is particulate matter small enough to pass through a size-selective inlet with a 50 % efficiency cut-off at x µm aerodynamic diameter — the measurand is thus defined by convention through the inlet design and operating conditions, since there are no reference materials for PM10 or PM2.5 (introduction of the standard).
PRINCIPLE (4.1). Air passes through a size-selective inlet at a known, constant flow rate, and the PM fraction is collected on a filter for nominally 24 h. The mass of PM is determined by weighing the filter before and after sampling at pre-specified, constant conditions. The result is affected by: the inlet design, the flow rate, deposition losses in the pipework between the inlet and the filter, volatilization of water and semi-volatile PM between sampling and weighing, and changes in filter mass (water, semi-volatile compounds, buoyancy, static electricity) — the standard sets requirements to minimize these effects.
QUALITY CONTROL (4.2). With each measurement: filter handling and conditioning, weighing room conditions, balance check and blank filters (clause 6 — weighing room blanks and field blanks). Less frequently: calibration of flow and balance, inlet cleaning and leak test of the system (clause 7). Measurement uncertainty is evaluated in accordance with clause 9 (GUM concept).
SAMPLERS (5.1.1). A newly procured reference sampler shall comply fully with the current edition. Type testing is performed by a competent body on two samplers of the same model in the laboratory and in the field, type approval is awarded by the competent authority of a Member State, and the type testing report shall be publicly available. Older samplers, including those described in Annex B of the 2014 edition, can still be used for monitoring and ongoing quality control if a justified additional allowance is made to their uncertainty.
REGULATIONS. In Poland the limit values for the protection of human health are: PM10 — 50 µg/m³ for 24 hours (35 exceedances permitted per year) and 40 µg/m³ for a year; PM2.5 — 20 µg/m³ for a year from 2020 (phase II; previously 25 µg/m³) — Dz.U. 2021 item 845, Annex 1 No 6–7. The reference methodology in the regulation on the assessment of substance levels is still PN-EN 12341:2014-07 (Dz.U. 2024 item 870, Annex 7 Table 1 No 7). The new Directive (EU) 2024/2881 on ambient air quality names EN 12341:2023 as the reference method, repeals Directive 2008/50/EC with effect from 12 December 2026, and Member States transpose it by 11 December 2026.
WHAT WE DO NOT GIVE HERE. Weighing room conditions, filter conditioning time, requirements for the balance and filter material, flow tolerances or inlet dimensions — they are in 5.1.5–7 and the annexes of the standard, outside the available text. The changes of the 2023 edition compared with 2014 (listed in Annex I).
Method principle
Ambient air flows at a constant nominal flow rate through a size-selective inlet as specified in Annex A of the standard, and the particulate matter passing the inlet is collected on a filter for nominally 24 h. The filter is weighed before and after sampling at pre-specified, constant conditions, and the mass gain divided by the air volume at ambient conditions near the inlet gives the concentration in µg/m³. The nominal flow rate is 2.3 m³/h; the PM10 or PM2.5 fraction is particulate matter passing an inlet with a 50 % efficiency cut-off at 10 or 2.5 µm aerodynamic diameter.
Applications
- Determination of PM10 or PM2.5 concentration in ambient air — rural areas, urban background, traffic-orientated locations and locations influenced by industrial sources (clause 1)
- Assessment of compliance with limit values: PM10 50 µg/m³ for a day and 40 µg/m³ for a year, PM2.5 20 µg/m³ for a year (Poland: Dz.U. 2021 item 845)
- Reference methodology for assessing levels of suspended particulate matter (Poland: Dz.U. 2024 item 870, Annex 7 — 2014 edition; Directive (EU) 2024/2881 — 2023 edition)
- Sampling of particulate matter for further determinations: heavy metals and PAH (EN 14902, EN 15549) and PM2.5 constituents (EN 16909, EN 16913) — introduction of the standard
- Demonstration of equivalence of candidate methods, including automated ones, in accordance with EN 16450 (clause 1, note 2)
- Type testing of samplers by a competent body (5.1.1)
Key parameters
| Parameter | Value |
|---|---|
| Flow rate | 2.3 m³/h — nominal (clause 1) |
| Sampling period | nominally 24 h (clause 1, 4.1) |
| Range of application (24 h) | from approximately 1 µg/m³ to 150 µg/m³ (PM10) and 120 µg/m³ (PM2.5) (clause 1) |
| Definition of PMx | particulate matter passing an inlet with a 50 % efficiency cut-off at x µm aerodynamic diameter (3.1.16) |
| Unit and conditions | µg/m³; air volume at ambient conditions near the inlet at the time of sampling (clause 1) |
| Blank filters | weighing room blanks and field blanks (3.1.9, 3.1.25, 6.7) |
| Uncertainty | evaluated in accordance with GUM (clause 9) |
Standard
- Standard number
- PN-EN 12341:2024-01
- Title (PL)
- Powietrze atmosferyczne — Standardowa grawimetryczna metoda pomiarowa oznaczania stężeń masowych frakcji PM10 lub PM2,5 pyłu zawieszonego
- Title (EN)
- Ambient air — Standard gravimetric measurement method for the determination of the PM10 or PM2,5 mass concentration of suspended particulate matter
Step-by-step procedure
-
Conditioning and weighing of filters before sampling
Condition and weigh the clean filters in the weighing room in accordance with 6.3 of the standard.
-
Loading cassettes
Load the filters into the sampler cassettes (6.4.1).
-
Sampling
Sample air through the size-selective inlet at a constant nominal flow rate of 2.3 m³/h for nominally 24 h per filter (clause 1, 4.1, 6.4.2).
-
Storage and transport
Store and transport the filters in accordance with 6.4.3, limiting the loss of water and semi-volatile constituents (4.1).
-
Conditioning and weighing after sampling
Condition and weigh the loaded filters in accordance with 6.5 and the weighing room procedures (6.6).
-
Blank filters
Run weighing room blanks and field blanks (6.7).
-
Calculation
Concentration = mass of particulate matter (difference between filter mass after and before sampling) divided by the air volume at ambient conditions near the inlet; result in µg/m³ (clause 1, clause 8).
-
Ongoing quality control
Check and calibrate the sensors, flow rate and balance, clean the inlet, perform the leak test and check the sampler clock (clause 7).
-
Uncertainty
Evaluate the uncertainty in accordance with clause 9 and compare it with the data quality objectives.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Sequential or single-filter sampler | meeting the requirements of the standard, type tested by a competent body (clause 1, 5.1.1) | — |
| PM10 or PM2.5 size-selective inlet | design as specified in Annex A of the standard, at a nominal flow rate of 2.3 m³/h | — |
| Flow control system, ambient temperature and pressure sensors | sampler components according to 5.1.6–5.1.8 of the standard | — |
| Weighing room and balance | conditions and requirements according to 6.2 of the standard | — |
| Flow meter for flow checks | checks and calibration of the sampler flow rate according to 7.7–7.8 | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Filters | — | filter material and suitability according to 5.1.5 of the standard; the standard gives criteria and test methods for the suitability of filters (introduction) |
Health and safety (OHS)
- The standard contains no safety requirements; the rules for working at the monitoring station and in the weighing room are set by the laboratory on the basis of a risk assessment
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 12341:2024-01 — “Ambient air — Standard gravimetric measurement method for the determination of the PM10 or PM2,5 mass concentration of suspended particulate matter”.
How does this method work?
Ambient air flows at a constant nominal flow rate through a size-selective inlet as specified in Annex A of the standard, and the particulate matter passing the inlet is collected on a filter for nominally 24 h. The filter is weighed before and after sampling at pre-specified, constant conditions, and the mass gain divided by the air volume at ambient conditions near the inlet gives the concentration in µg/m³.
What is the measuring range and accuracy?
Flow rate: 2.3 m³/h — nominal (clause 1); Sampling period: nominally 24 h (clause 1, 4.1); Range of application (24 h): from approximately 1 µg/m³ to 150 µg/m³ (PM10) and 120 µg/m³ (PM2.5) (clause 1); Definition of PMx: particulate matter passing an inlet with a 50 % efficiency cut-off at x µm aerodynamic diameter (3.1.16).
How long does the test take?
The procedure comprises 9 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Sequential or single-filter sampler, PM10 or PM2.5 size-selective inlet, Flow control system, ambient temperature and pressure sensors, Weighing room and balance, Flow meter for flow checks.
Where is this test used?
Determination of PM10 or PM2.5 concentration in ambient air — rural areas, urban background, traffic-orientated locations and locations influenced by industrial sources (clause 1); Assessment of compliance with limit values: PM10 50 µg/m³ for a day and 40 µg/m³ for a year, PM2.5 20 µg/m³ for a year (Poland: Dz.U. 2021 item 845); Reference methodology for assessing levels of suspended particulate matter (Poland: Dz.U. 2024 item 870, Annex 7 — 2014 edition; Directive (EU) 2024/2881 — 2023 edition); Sampling of particulate matter for further determinations: heavy metals and PAH (EN 14902, EN 15549) and PM2.5 constituents (EN 16909, EN 16913) — introduction of the standard; Demonstration of equivalence of candidate methods, including automated ones, in accordance with EN 16450 (clause 1, note 2); Type testing of samplers by a competent body (5.1.1).
What safety precautions apply?
The standard contains no safety requirements; the rules for working at the monitoring station and in the weighing room are set by the laboratory on the basis of a risk assessment.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 12341.