⚡ Water Electrical Conductivity
In short
Measurement of the electrical conductivity of all types of water with a conductivity meter fitted with a cell of known cell constant. The test is performed according to PN-EN 27888:1999; Reference temperature: 25.0 °C; precise measurement at 25.0 ± 0.1 °C. The procedure comprises 6 steps; it is used for: Water intended for human consumption — electrical conductivity 2500 µS/cm, determined at 25 °C (Journal of Laws 2026, item 748, Annex 1 Part C); a group A monitoring parameter (Annex 3), Monitoring the quality of surface waters (clause 1), Process waters in water supply and treatment plants (clause 1).
At a glance
- Standard: PN-EN 27888:1999
- Category: Physicochemistry
- Procedure steps: 6
- Reference temperature: 25.0 °C; precise measurement at 25.0 ± 0.1 °C
- Result: γ25 in mS/m or other units (1 mS/m = 10 µS/cm)
- KCl standards at 25 °C: 0.1 mol/l — 1290 mS/m; 0.01 mol/l — 141 mS/m; 0.001 mol/l — 14.7 mS/m
Overview
STATUS (as of 28.10.2026). PN-EN 27888:1999 — Polish version, published 27.05.1999, implements EN 27888:1993 [IDT] and ISO 7888:1985 [IDT]; it replaced PN-C-04542:1977. The standard is current — no withdrawal note.
SCOPE (clause 1). The standard specifies a method for measuring the electrical conductivity of all types of water. Conductivity is used to monitor the quality of surface waters, process waters in water supply and treatment plants, and waste waters; it also makes it possible to check the completeness of an analysis of ionic constituents. Sometimes absolute values are important, sometimes only relative changes.
TERMS (clauses 2–3). Electrical conductivity (γ) is the reciprocal of the resistance, measured under specified conditions, between the opposite faces of a unit cube of an aqueous solution; it is expressed in S/m (1 S/m = 10⁴ µS/cm = 10³ mS/m). It is a measure of the current conducted by the ions and depends on their concentration and nature, on the temperature and on the viscosity of the solution. Even pure water has a small conductivity as a result of its own dissociation.
TEMPERATURE (2.4, 7.2). Results are compared after conversion to the reference temperature of 25.0 °C. Measurement at 25.0 ± 0.1 °C is the most accurate; a measurement at another temperature is corrected with the instrument's temperature compensator (calibrated strictly according to the manufacturer), with the temperature coefficient of the sample — γ25 = γθ / [1 + (α/100)(θ − 25)] — or with the f25 factors of Table 3; each of these corrections gives a less accurate result than a measurement at 25.0 °C. The factors of Table 3 apply to natural waters with a γ25 of about 6–100 mS/m and are not applicable to potassium chloride solutions — therefore calibration is always carried out at 25.0 ± 0.1 °C. Earlier results quoted as γ20 can be converted to 25 °C for such waters with the factor 1.116.
INTERFERENCES (clause 9). The result is affected by contamination of the sample in the cell; suspended matter, grease and oil can foul the electrodes and change the cell constant, which can only be detected by checking with a potassium chloride standard solution. Air bubbles collecting on the electrodes when the sample is warmed to 25 °C interfere with the measurement. Conductivities below 1 mS/m are influenced by carbon dioxide and ammonia from the air — such waters are measured in a flow-type cell; for water of very low ion content the contribution of the conductivity of the water itself may be significant.
DRINKING WATER — LEGAL REQUIREMENTS. Regulation of the Minister of Health of 22 May 2026 (Journal of Laws 2026, item 748): electrical conductivity 2500 µS/cm (250 mS/m), determined at 25 °C; the water should not be corrosive (Annex 1 Part C Table 2 item 8). Conductivity is one of the group A monitoring parameters (Annex 3); measurement uncertainty — 20 % of the parametric value (Annex 4 Part B).
WATER MONITORING — LEGAL REQUIREMENTS. Regulation of the Minister of Infrastructure of 13 July 2021 (Journal of Laws 2021, item 1576), Annex 8, indicator 3.3.2: in the monitoring of surface water bodies and groundwater bodies, the reference methodology for the indicator “specific electrolytic conductivity at 20 °C” (electrometric, conductometric method) is PN-EN 27888 or a test procedure; another methodology may be used once its equivalence with the reference methodology has been demonstrated (explanatory note 1).
WHAT WE DO NOT GIVE HERE. The values of Table 2 of the standard (recommended cell constants for the measuring ranges) and the standard deviations of Tables 4 and 5 (measurements by one laboratory in the United Kingdom). Repeatability depends on the instrument — the standard refers to the manufacturer's data (8.2).
Method principle
Direct measurement of the electrical conductivity of an aqueous solution with a suitable instrument. Conductivity is a measure of the current conducted by the ions present in the water and depends on the concentration of the ions, their nature, the temperature and the viscosity of the solution.
The electrical conductivity is obtained by multiplying the measured conductance by the cell constant K = l/A (m⁻¹), which results from the geometry of the cell and is determined empirically with potassium chloride standard solutions. Many instruments have a built-in cell constant correction and display the electrical conductivity directly.
Applications
- Water intended for human consumption — electrical conductivity 2500 µS/cm, determined at 25 °C (Journal of Laws 2026, item 748, Annex 1 Part C); a group A monitoring parameter (Annex 3)
- Monitoring the quality of surface waters (clause 1)
- Process waters in water supply and treatment plants (clause 1)
- Waste waters (clause 1)
- Checking the completeness of an analysis of the ionic constituents of water (clause 1)
Key parameters
| Parameter | Value |
|---|---|
| Reference temperature | 25.0 °C; precise measurement at 25.0 ± 0.1 °C |
| Result | γ25 in mS/m or other units (1 mS/m = 10 µS/cm) |
| KCl standards at 25 °C | 0.1 mol/l — 1290 mS/m; 0.01 mol/l — 141 mS/m; 0.001 mol/l — 14.7 mS/m |
| Water for solutions | double-distilled or deionized, γ25 < 0.1 mS/m |
| Cell constant | checked at least once every 6 months |
| Conductivity below 1 mS/m | measurement in a flow-type cell from which air is excluded |
Standard
- Standard number
- PN-EN 27888:1999
- Title (PL)
- Jakość wody — Oznaczanie przewodności elektrycznej właściwej
- Title (EN)
- Water quality — Determination of electrical conductivity
Step-by-step procedure
-
Sampling
Into a polyethylene bottle, completely filled and tightly stoppered; not into soda glass bottles. The measurement is made as soon as practicable, particularly when gases (carbon dioxide, ammonia) may be exchanged with the air or biological activity is possible (clause 6).
-
Storage
Biological activity is reduced by storing in the dark at 4 °C; before the measurement the sample is brought to equilibrium at 25 °C. No suitable preservative is known (clause 6).
-
Instrument and cell constant
Prepare the instrument according to the manufacturer, with a cell of known cell constant appropriate to the range (Table 2). If the cell constant is not accurately known — determine it with potassium chloride standard solutions appropriate to the range; check the cell constant at least once every 6 months (7.1).
-
Measurement
For high-precision work, measure when the sample and the parts of the apparatus in contact with it have reached 25.0 ± 0.1 °C — this eliminates errors from compensation and conversion. Instrument without cell constant correction: electrical conductivity = measured conductance × cell constant (7.1).
-
Temperature correction
When measurement at 25.0 °C is not possible (e.g. in the field), measure at a known temperature θ and convert with the instrument's compensator, the temperature coefficient of the sample or the f25 factor of Table 3 for natural waters (7.2).
-
Result
The result is read from the instrument and reported as γ25 in mS/m with reference to the standard; for a measurement at another temperature the measuring temperature and the method of correction are also reported, e.g. “γ25 = 25.8 mS/m (according to ISO 7888), measuring temperature 11.5 °C, mathematical correction” (clause 8).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Conductivity meter | with a flow-type or dip-type cell with two or more electrodes, or with induction-type electrodes; preferably for discrete and continuous measurement, in the laboratory and in the field (5.1) | — |
| Electrodes | platinum electrodes for precise measurements are platinized; unplatinized electrodes are used only for field work and routine laboratory testing. When not in use, the cell is filled with water so that the electrodes do not dry out (5.2) | — |
| Thermometer | accuracy ± 0.1 °C for precise determinations, ± 0.5 °C for routine ones (5.3) | — |
| Thermostatic bath | 25.0 ± 0.1 °C; ± 0.5 °C is sufficient for routine measurements (5.4) | — |
| Polyethylene bottles | for samples, completely filled and tightly stoppered; soda glass bottles are not used (clause 6) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Water for solutions | — | double-distilled or deionized, γ25 < 0.1 mS/m (4.1) |
| Potassium chloride — standard solution A, 0.1 mol/l | 7447-40-7 | dissolve 7.456 g of KCl dried at 105 °C for 2 h in water and dilute to 1000 ml; γ25 = 1290 mS/m (4.2) |
| Potassium chloride — standard solution B, 0.01 mol/l | 7447-40-7 | dilute 100 ml of solution A with water to 1000 ml; γ25 = 141 mS/m (4.3) |
| Potassium chloride — standard solution C, 0.001 mol/l | 7447-40-7 | dilute 100 ml of solution B to 1000 ml with water freed from carbon dioxide (purging with pure nitrogen or boiling), shortly before use and with minimal contact with air; γ25 = 14.7 mS/m (4.4) |
| Platinizing solution | — | 1.5 g of hydrogen hexachloroplatinate(IV) hexahydrate in 50 ml of water containing 0.0125 g of lead(II) acetate (4.5) |
Health and safety (OHS)
- The standard gives no safety warnings; reagents, including the platinizing solution (4.5), are handled according to the safety data sheets
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 27888:1999 — “Water quality — Determination of electrical conductivity”.
How does this method work?
Direct measurement of the electrical conductivity of an aqueous solution with a suitable instrument. Conductivity is a measure of the current conducted by the ions present in the water and depends on the concentration of the ions, their nature, the temperature and the viscosity of the solution.
What is the measuring range and accuracy?
Reference temperature: 25.0 °C; precise measurement at 25.0 ± 0.1 °C; Result: γ25 in mS/m or other units (1 mS/m = 10 µS/cm); KCl standards at 25 °C: 0.1 mol/l — 1290 mS/m; 0.01 mol/l — 141 mS/m; 0.001 mol/l — 14.7 mS/m; Water for solutions: double-distilled or deionized, γ25 < 0.1 mS/m.
How long does the test take?
The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Conductivity meter, Electrodes, Thermometer, Thermostatic bath, Polyethylene bottles.
Where is this test used?
Water intended for human consumption — electrical conductivity 2500 µS/cm, determined at 25 °C (Journal of Laws 2026, item 748, Annex 1 Part C); a group A monitoring parameter (Annex 3); Monitoring the quality of surface waters (clause 1); Process waters in water supply and treatment plants (clause 1); Waste waters (clause 1); Checking the completeness of an analysis of the ionic constituents of water (clause 1).
What safety precautions apply?
The standard gives no safety warnings; reagents, including the platinizing solution (4.5), are handled according to the safety data sheets.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 27888.