⚗️ Trace Elements in Food — Pressure Digestion of Samples for the Determination of Metals
In short
Pressure digestion of foodstuffs before the determination of elements: the sample is digested with nitric acid (sometimes with other acids or hydrogen peroxide added) in a closed vessel at a temperature of at least 180 °C, with conventional or microwave heating. The test is performed according to PN-EN 13805:2014-11; Digestion temperature: at least 180 °C; above 230 °C quartz inserts are advantageous. The procedure comprises 6 steps; it is used for: Foodstuffs — preparation of a solution for the determination of elements, including metals (clause 1), Combination with flame, electrothermal, hydride and cold-vapour AAS and with ICP-MS — tested in interlaboratory studies; ICP-OES, voltammetry, atomic fluorescence — possible (clause 1), Determination of tin, antimony and iron — digestion with hydrochloric acid added; foods containing silicates or titanium dioxide — with hydrofluoric acid (6.3.2).
At a glance
- Standard: PN-EN 13805:2014-11
- Category: Food Chemistry
- Procedure steps: 6
- Digestion temperature: at least 180 °C; above 230 °C quartz inserts are advantageous
- Sample mass: carbon-rich samples usually up to 200 mg of fresh mass; leaner ones up to 3 g or 4 ml of liquid; constant ratio of sample to acid
- Acids: nitric acid at least 65 %, usually 3 ml; H₂O₂ from 0.5 ml to 1 ml; HCl at least 0.5 ml for Sn, Sb, Fe; HF for silicates and TiO₂
Overview
STATUS (as of 02.11.2026). PN-EN 13805:2014-11 — English version with a Polish title (in translation: “Foodstuffs — Determination of trace elements — Pressure digestion”), published on 28.11.2014, implements EN 13805:2014 [IDT]; replaced PN-EN 13805:2003 (the Polish version); the PKN catalogue card has no withdrawal note (read on 02.11.2026). EN 13805:2014 replaced EN 13805:2002; changes according to the foreword: restructuring of some clauses, new subclauses on apparatus with conventional heating, with microwave heating and on the microwave autoclave (5.2.2–5.2.4), a revised subclause on the addition of acid (6.3.2), completely new digestion examples (6.4), a supplemented bibliography and editorial revision of the whole text. The description is based on the preview of EN 13805:2014 (iTeh, SIST EN 13805:2015: clauses 1–5 and 6.1–6.3.3).
SCOPE AND PRINCIPLE (clauses 1 and 3). A method for the pressure digestion of foodstuffs intended for the determination of elements. It has been tested in interlaboratory studies in combination with atomic absorption spectrometry (flame, electrothermal, hydride, cold vapour) and ICP-MS; it can be combined with other techniques, e.g. ICP-OES, voltammetry or atomic fluorescence. The physicochemical digestion is intended to decompose the sample material and give a test solution containing the elements to be determined; it is used when the measurement has been validated in combination with this digestion. The result refers to the total content of the element, depending on the reagents and the measurement procedure used. The sample is homogenized without introducing contamination and then digested with nitric acid (sometimes with other acids added) at high temperature and under pressure in a closed vessel, with conventional or microwave heating. General requirements for the determination of elements in foodstuffs are given in EN 13804 (normative reference).
REAGENTS (clause 4). The trace element content of the reagents and water shall be low enough not to affect the result; nitric and hydrochloric acid of insufficient purity are purified by sub-boiling distillation. Nitric acid — mass fraction of at least 65 %, density about 1.4 g/ml; diluted nitric acid — at least (1 + 9); hydrochloric acid — at least 30 %, about 1.15 g/ml; hydrofluoric acid — at least 40 %, about 1.14 g/ml; hydrogen peroxide — at least 30 %, about 1.11 g/ml.
APPARATUS (clause 5). Everything that comes into contact with the sample is washed with diluted nitric acid and water; vessels can be cleaned with hot acid vapour in an apparatus intended for this purpose. Pressure vessels shall be commercially available and safety-tested, with acid-resistant inserts made of materials with a low risk of contamination, withstand at least 180 °C and have safety devices (e.g. rupture discs, a pressure jacket, power control). With conventional heating, inserts of PTFE, quartz glass, FEP or PFA are used. Microwave systems shall have temperature measurement (internal or external; an infrared sensor only if the manufacturer confirms that the reading corresponds to the temperature of the solution) — without it the system shall not be used; inserts of PTFE, PFA, TFM or quartz. In a microwave autoclave the vessels are closed by a pre-pressure of at least 40 bar. Above 230 °C quartz glass is advantageous; for the determination of mercury and of elements that are readily adsorbed on rough surfaces (e.g. antimony), quartz is recommended. In addition: a heatable ultrasonic bath and a sub-boiling distillation apparatus for acids made of quartz or fluoropolymers.
PROCEDURE (clauses 6.1–6.3.3). Contamination at every stage shall be kept as low as possible; samples are prepared as in a household, and for the determination of chromium and nickel no stainless steel equipment is used. The sample mass is chosen to suit the capacity of the vessel and the permissible pressure increase, without departing from the manufacturer’s data: for carbon-rich samples (fats, carbohydrate-rich products) usually not more than 200 mg of fresh mass (to the nearest 1 mg), with a lower carbon content up to 3 g of fresh mass or 4 ml of liquid; when the mass is changed, the ratio of sample mass to acid volume is kept. 1 ml of water is added to a powder (about 200 mg); none is added to fatty materials. Alcohol is removed from alcohol-containing samples before the acid is added, by gentle heating, e.g. in an ultrasonic bath at 60 °C. Usually 3 ml of nitric acid are sufficient (microwave systems usually require more); 0.5 ml to 1 ml of hydrogen peroxide may be added, which prevents the sample from adhering to the walls, improves mixing and reduces the amount of nitrous gases. The sample shall be covered with acid before the vessel is closed; a pre-reaction at room temperature in an open, loosely covered vessel is recommended (reactive samples — overnight). For the determination of tin, antimony and iron at least 0.5 ml of hydrochloric acid is added — only after the reaction with nitric acid has ended — and the vessel is closed immediately. Foods containing silicates or titanium dioxide may require hydrofluoric acid — then only inserts made of fluoroplastics, because quartz is irreversibly damaged. Digestion takes place at a temperature of at least 180 °C; in microwave systems heating is slow, at reduced power. The required temperature — and thus the degree of decomposition — is set according to the subsequent measurement procedure: a higher temperature gives less residual carbon in the solution and less background interference in ET-AAS and ICP-OES.
SECOND SOURCE. The FDA method from the Elemental Analysis Manual (EAM 4.7, version 1.3 of April 2025) combines microwave-assisted digestion with nitric acid with ICP-MS determination of, among others, arsenic, cadmium, chromium, lead and mercury in food; it requires a digestion system with temperature control up to at least 200 °C and opens the vessels after cooling below 50 °C.
REGULATIONS. Maximum levels of lead, cadmium, mercury, arsenic and inorganic tin in food are laid down in Regulation (EU) 2023/915 (Annex I, Section 3 “Metals and other elements”), which repealed Regulation (EC) No 1881/2006; the values depend on the type of food.
WHAT WE DO NOT GIVE HERE. The digestion time, cooling and further treatment of the solution, quality control and example digestion programmes (6.3.4–6.4) — these subclauses are not in the preview of the standard. Nor do we give ICP-MS measuring conditions or limits of quantification — the standard does not cover them.
Method principle
A food sample homogenized without contamination is digested with nitric acid — sometimes with hydrogen peroxide, hydrochloric acid or hydrofluoric acid added — in a closed pressure vessel at a temperature of at least 180 °C, with conventional or microwave heating. The organic matter is decomposed and the elements to be determined pass into the test solution.
The solution is analysed by a technique validated together with this digestion — in the interlaboratory study these were flame, electrothermal, hydride and cold-vapour AAS and ICP-MS. The result refers to the total content of the element.
Applications
- Foodstuffs — preparation of a solution for the determination of elements, including metals (clause 1)
- Combination with flame, electrothermal, hydride and cold-vapour AAS and with ICP-MS — tested in interlaboratory studies; ICP-OES, voltammetry, atomic fluorescence — possible (clause 1)
- Determination of tin, antimony and iron — digestion with hydrochloric acid added; foods containing silicates or titanium dioxide — with hydrofluoric acid (6.3.2)
- Checking compliance with the maximum levels of lead, cadmium, mercury, arsenic and tin in Regulation (EU) 2023/915 — after determination by a validated measurement technique
Key parameters
| Parameter | Value |
|---|---|
| Digestion temperature | at least 180 °C; above 230 °C quartz inserts are advantageous |
| Sample mass | carbon-rich samples usually up to 200 mg of fresh mass; leaner ones up to 3 g or 4 ml of liquid; constant ratio of sample to acid |
| Acids | nitric acid at least 65 %, usually 3 ml; H₂O₂ from 0.5 ml to 1 ml; HCl at least 0.5 ml for Sn, Sb, Fe; HF for silicates and TiO₂ |
| Microwave system | temperature measurement mandatory; autoclave closed by a pre-pressure of at least 40 bar |
| Prohibitions | no perchloric acid; remove alcohol before adding acid |
| Measurement techniques | AAS (flame, electrothermal, hydride, cold vapour) and ICP-MS — tested; ICP-OES, voltammetry, AFS — possible |
Standard
- Standard number
- PN-EN 13805:2014-11
- Title (PL)
- Artykuły żywnościowe — Oznaczanie pierwiastków śladowych — Mineralizacja ciśnieniowa
- Title (EN)
- Foodstuffs — Determination of trace elements — Pressure digestion
Step-by-step procedure
-
Preparation of the sample
Homogenization without contamination; for Cr and Ni without stainless steel equipment; in accordance with EN 13804 (6.2).
-
Test portion
Mass chosen to suit the vessel and the permissible pressure: carbon-rich samples usually up to 200 mg (to the nearest 1 mg), leaner ones up to 3 g or 4 ml; powder with 1 ml of water; alcohol removed at 60 °C (6.3.1).
-
Addition of acid and pre-reaction
Usually 3 ml of nitric acid, optionally 0.5 ml to 1 ml of H₂O₂; sample completely covered with acid; pre-reaction in an open, loosely covered vessel, for reactive samples overnight (6.3.2).
-
Additions for selected elements
Sn, Sb, Fe — at least 0.5 ml of HCl after the reaction with nitric acid has ended and immediate closing of the vessel; silicates, TiO₂ — HF only in inserts made of fluoroplastics (6.3.2).
-
Digestion
At a temperature of at least 180 °C; in a microwave system slow heating at reduced power; the temperature is chosen to suit the measurement procedure (6.3.3).
-
Test solution
Further treatment and determination according to the standard validated with this digestion (clause 3; details in 6.3.4–6.4 outside the preview).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Pressure digestion apparatus with conventional heating | with or without an external autoclave; inserts of PTFE, quartz glass, FEP or PFA (5.2.2) | — |
| Digestion apparatus with microwave heating | measurement of the solution temperature mandatory; inserts of PTFE, PFA, TFM or quartz (5.2.3) | — |
| Microwave autoclave | vessels closed by a pre-pressure of at least 40 bar; vessels of fluoroplastics or quartz (5.2.4) | — |
| Heatable ultrasonic bath | e.g. for removing alcohol from the sample at 60 °C before adding acid (5.3, 6.3.1) | — |
| Sub-boiling distillation apparatus for acids and apparatus for cleaning vessels with acid vapour | made of quartz or fluoropolymers — purification of nitric and hydrochloric acid, cleaning of vessels (5.4–5.5) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Nitric acid | 7697-37-2 | mass fraction of at least 65 %, about 1.4 g/ml; usually 3 ml per digestion; diluted (1 + 9) for washing vessels (4.1–4.2, 6.3.2) |
| Hydrogen peroxide | 7722-84-1 | at least 30 %, about 1.11 g/ml; from 0.5 ml to 1 ml (4.5, 6.3.2) |
| Hydrochloric acid | 7647-01-0 | at least 30 %, about 1.15 g/ml; at least 0.5 ml for the determination of Sn, Sb and Fe (4.3, 6.3.2) |
| Hydrofluoric acid | 7664-39-3 | at least 40 %, about 1.14 g/ml; for foods containing silicates or TiO₂, only in inserts made of fluoroplastics (4.4, 6.3.2) |
Health and safety (OHS)
- Perchloric acid shall not be used for pressure digestion, not even in small amounts; digestion of carbon-rich materials (cellulose, carbohydrates, fats, oils) can lead to an explosion (6.1, 6.3.2)
- Alcohol must be evaporated before nitric acid is added — it reacts with it violently after a delay (6.1, 6.3.1)
- A sample not immersed in the acid can locally overheat the vessel in a microwave system, up to its rupture (warning in 6.3.2)
- Beware of nitrous gases; before using the apparatus read the manufacturer’s instructions and safety rules, and when using hydrofluoric acid follow the appropriate safety guidelines (6.1, 6.3.2)
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 13805:2014-11 — “Foodstuffs — Determination of trace elements — Pressure digestion”.
How does this method work?
A food sample homogenized without contamination is digested with nitric acid — sometimes with hydrogen peroxide, hydrochloric acid or hydrofluoric acid added — in a closed pressure vessel at a temperature of at least 180 °C, with conventional or microwave heating. The organic matter is decomposed and the elements to be determined pass into the test solution.
What is the measuring range and accuracy?
Digestion temperature: at least 180 °C; above 230 °C quartz inserts are advantageous; Sample mass: carbon-rich samples usually up to 200 mg of fresh mass; leaner ones up to 3 g or 4 ml of liquid; constant ratio of sample to acid; Acids: nitric acid at least 65 %, usually 3 ml; H₂O₂ from 0.5 ml to 1 ml; HCl at least 0.5 ml for Sn, Sb, Fe; HF for silicates and TiO₂; Microwave system: temperature measurement mandatory; autoclave closed by a pre-pressure of at least 40 bar.
How long does the test take?
The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Pressure digestion apparatus with conventional heating, Digestion apparatus with microwave heating, Microwave autoclave, Heatable ultrasonic bath, Sub-boiling distillation apparatus for acids and apparatus for cleaning vessels with acid vapour.
Where is this test used?
Foodstuffs — preparation of a solution for the determination of elements, including metals (clause 1); Combination with flame, electrothermal, hydride and cold-vapour AAS and with ICP-MS — tested in interlaboratory studies; ICP-OES, voltammetry, atomic fluorescence — possible (clause 1); Determination of tin, antimony and iron — digestion with hydrochloric acid added; foods containing silicates or titanium dioxide — with hydrofluoric acid (6.3.2); Checking compliance with the maximum levels of lead, cadmium, mercury, arsenic and tin in Regulation (EU) 2023/915 — after determination by a validated measurement technique.
What safety precautions apply?
Perchloric acid shall not be used for pressure digestion, not even in small amounts; digestion of carbon-rich materials (cellulose, carbohydrates, fats, oils) can lead to an explosion (6.1, 6.3.2); Alcohol must be evaporated before nitric acid is added — it reacts with it violently after a delay (6.1, 6.3.1); A sample not immersed in the acid can locally overheat the vessel in a microwave system, up to its rupture (warning in 6.3.2); Beware of nitrous gases; before using the apparatus read the manufacturer’s instructions and safety rules, and when using hydrofluoric acid follow the appropriate safety guidelines (6.1, 6.3.2).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 13805.