🔬 Manganese in water
In short
Spectrometric determination of total manganese (dissolved, suspended and organically bound) in surface water and drinking water with formaldoxime. The test is performed according to ISO 6333:1986; Range: 0.01–5 mg/l; higher concentrations after dilution (clause 1). The procedure comprises 8 steps; it is used for: Water intended for human consumption — manganese 50 µg/l (Journal of Laws 2026, item 748), Surface waters (clause 1), Total manganese, including suspended and organically bound manganese — after pretreatment (6.2).
At a glance
- Standard: ISO 6333:1986
- Category: Physicochemistry
- Procedure steps: 8
- Range: 0.01–5 mg/l; higher concentrations after dilution (clause 1)
- Wavelength: about 450 nm; measured against water
- Cells: 100 mm for the range 0–0.5 mg/l, 10 mm for the range 0–5 mg/l
Overview
STATUS (as of 29.10.2026). ISO 6333:1986 — current in the ISO catalogue, with no newer edition or draft. The standard has not been adopted as a Polish Standard — the catalogue of Polish Standards contains no standard numbered ISO 6333; the Polish title is a working translation. The regulations on waste water (Journal of Laws 2019, item 1311) and on water monitoring (Journal of Laws 2021, item 1576) do not refer to this standard.
SCOPE (clause 1). Total manganese — dissolved, suspended and organically bound — in surface water and drinking water, at concentrations of 0.01–5 mg/l; higher concentrations after suitable dilution of the sample. The method is not applicable to highly contaminated waters such as industrial waste water.
MANGANESE IN WATER. In water containing oxygen most of the manganese is present in undissolved forms, often associated with microorganisms, and in complexes, e.g. with humic acid; in oxygen-free or strongly acidic water all manganese is dissolved (introduction of the standard). Manganese(II) compounds predominate in ground waters; 0.01–4.08 g Mn/m³ has been found in ground waters in Poland. Manganese in tap water does not endanger health, but causes black deposits and stains.
INTERFERENCES (clause 8). Iron(II) forms a violet complex with formaldoxime. Adding EDTA and the hydroxylammonium chloride/ammonia solution reduces this interference, but the best way is to add a constant, known amount of iron(II) (as ammonium iron(II) sulfate) to every calibration solution, blank and test solution. 1 mg/l of cobalt gives a response equivalent to 40 µg/l of manganese. In the presence of calcium, orthophosphate above 2 mg P/l causes low results, and calcium plus magnesium above 300 mg/l in total cause high results. Turbidity after formation of the complex is removed by centrifuging before the absorbance is measured.
PRECISION (7.2, interlaboratory trial of 1982, 30 results each). E.g. 0.050 mg/l — mean 0.049 mg/l, standard deviation 0.0035 mg/l; 1.000 mg/l — 1.01 and 0.04 mg/l (laboratory A) and 1.001 and 0.0091 mg/l (laboratory B); 4.000 mg/l — 4.02 and 0.047 mg/l (A) and 4.198 and 0.1333 mg/l (B).
LEGAL REQUIREMENTS. Water intended for human consumption: manganese 50 µg/l (Regulation of the Minister of Health of 22 May 2026, Journal of Laws 2026, item 748, Annex 1 Part C Table 2 item 5); measurement uncertainty up to 30 % of the parametric value (Annex 4 Part B).
Method principle
A formaldoxime solution is added to the test portion; manganese forms an orange-red complex whose absorbance is measured at about 450 nm — at the absorption maximum (molar absorption coefficient 11 × 10³ l/(mol·cm)). The complex is stable between pH 9.5 and 10.5, the colour intensity is proportional to the amount of manganese, and the relationship between absorbance and concentration is linear up to 5 mg/l. Suspended or organically bound manganese requires a pretreatment that converts it into forms capable of reacting with formaldoxime — oxidation with peroxodisulfate (clause 2, 6.2).
Applications
- Water intended for human consumption — manganese 50 µg/l (Journal of Laws 2026, item 748)
- Surface waters (clause 1)
- Total manganese, including suspended and organically bound manganese — after pretreatment (6.2)
- Not for highly contaminated waters such as industrial waste water (clause 1)
Key parameters
| Parameter | Value |
|---|---|
| Range | 0.01–5 mg/l; higher concentrations after dilution (clause 1) |
| Wavelength | about 450 nm; measured against water |
| Cells | 100 mm for the range 0–0.5 mg/l, 10 mm for the range 0–5 mg/l |
| pH range of complex stability | 9.5–10.5 |
| Time | 5–10 min after adding NaOH, then at least 1 h after the hydroxylammonium chloride/ammonia solution; measurement 1–4 h after colour development |
| Test portion | 50 ml of sample acidified to pH about 1 (not lower), containing less than 0.25 mg of manganese |
Standard
- Standard number
- ISO 6333:1986
- Title (PL)
- Jakość wody — Oznaczanie manganu — Metoda spektrometryczna z formaldoksymem
- Title (EN)
- Water quality — Determination of manganese — Formaldoxime spectrometric method
Step-by-step procedure
-
Sampling and preservation
Container of polyethylene, PVC or glass; acidify the sample with 3 mol/l sulfuric acid to pH approximately, but not less than, 1 — this minimizes adsorption of manganese on the container walls and assists the dissolution of colloidal and particulate forms (clause 5).
-
Test portion
50 ml of the acidified sample, or an accurately measured smaller aliquot diluted to 50 ml, containing less than 0.25 mg of manganese (6.1).
-
Pretreatment
If organically bound or suspended manganese is present, add (225 ± 25) mg of oxidizing reagent to the test portion and: a) autoclave for 30 min in a bottle, cool and add about 0.5 g of sodium sulfite, or b) boil for about 40 min in a 100 ml conical flask or beaker, cool, transfer to a 50 ml volumetric flask, make up to the mark and add about 0.5 g of sodium sulfite. Autoclaving is preferable for samples containing humic acids; the pretreated sample may be kept overnight. Pretreatment destroys turbidity and colour; if experience shows that it is not required (e.g. in most cases for drinking water), it may be omitted (6.2).
-
Colour development
Add 1 ml of ammonium iron(II) sulfate solution and 2 ml of EDTA solution, mix, add 1 ml of formaldoxime solution and immediately 2 ml of sodium hydroxide solution (2.5 ml if the sample has been pretreated). Mix thoroughly, allow to stand for 5–10 min, then add, while mixing, 3 ml of the hydroxylammonium chloride/ammonia solution and leave to stand for at least 1 h (6.4.2, 6.5.1).
-
Measurement
Between 1 and 4 h after colour development, measure the absorbance at 450 nm against water; cells of 100 mm for the range 0–0.5 mg/l, 10 mm for the range 0–5 mg/l (6.4.3, 6.5.2).
-
Blank test
In parallel, with 50 ml of water instead of the test portion; if its absorbance differs significantly from the extrapolated absorbance of the zero member, the reasons must be investigated (6.3).
-
Calibration
Range A (0–0.5 mg/l): dilute 5 ml of standard solution to 1000 ml and put 0, 10, 20, 30 and 40 ml of it into 50 ml flasks — 0; 0.1; 0.2; 0.3 and 0.4 mg/l. Range B (0–5 mg/l): 50 ml of standard solution to 1000 ml and the same volumes — 0, 1, 2, 3 and 4 mg/l. Colour is developed as in step 4. The graph must be linear; the calibration factor f is the reciprocal of the slope (it may also be calculated by regression), and the intercept on the ordinate is the extrapolated absorbance of the zero member. The graph is checked periodically, especially when new reagents are used (6.4).
-
Calculation
ρMn = f · (A₁ − A₀) · g mg/l, where A₁ is the absorbance of the test solution, A₀ the extrapolated absorbance of the zero member and g = V₁ / V₂ (50 ml / volume of the aliquot taken); the volume of acid added to the sample is taken into account. Results to the nearest 0.01 mg/l in the range 0.01–1 mg/l and to 0.1 mg/l above 1 mg/l (clause 7).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Spectrometer | with continuous (prism, grating) or discontinuous (narrow-band filter) wavelength selection, for measurement at about 450 nm; cells with optical path lengths up to 100 mm (concentrations below 0.3 mg/l) and 10 mm (above 0.3 mg/l) (4.1) | — |
| Glass bottles, 100 ml | with ground-glass stoppers and metal clamps or with screw caps of non-coloured plastic, suitable for autoclaving (4.2) | — |
| Autoclave or pressure cooker | capable of maintaining 120 °C and 200 kPa (4.3) | — |
| Glassware and containers | washed with approximately 1 mol/l hydrochloric acid and rinsed with water; sample containers of polyethylene, PVC or glass (clauses 4–5) | — |
| 50 ml and 1000 ml volumetric flasks, pipettes | for the test portion and calibration solutions (6.4) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Oxidizing reagent | — | potassium peroxodisulfate K₂S₂O₈ or sodium peroxodisulfate Na₂S₂O₈ (3.1) |
| Sodium sulfite, anhydrous | 7757-83-7 | about 0.5 g after oxidation — reduces oxidizing substances (3.2) |
| EDTA (tetrasodium salt) solution, c = 0.24 mol/l | — | dissolve 90 g of disodium EDTA dihydrate and 19 g of NaOH in water and dilute to 1000 ml; alternatively 109 g of tetrasodium EDTA tetrahydrate or 100 g of the dihydrate (3.3) |
| Formaldoxime solution | — | dissolve 10 g of hydroxylammonium chloride in about 50 ml of water, add 5 ml of 35 % (m/m) formaldehyde solution (ρ = 1.08 g/ml) and dilute to 100 ml; keep the bottle in a dark, cool place; shelf-life at least one month (3.4) |
| Hydroxylammonium chloride/ammonia solution | — | equal volumes of 6 mol/l hydroxylammonium chloride solution (42 g to 100 ml) and 4.7 mol/l ammonia solution (70 ml of concentrated ammonia, ρ = 0.91 g/ml, to 200 ml) (3.5) |
| Ammonium iron(II) sulfate hexahydrate solution, 700 mg/l | — | dissolve 700 mg in water, add 1 ml of 3 mol/l sulfuric acid and dilute to 1000 ml (3.6) |
| Sulfuric acid, c(H₂SO₄) = 3 mol/l | 7664-93-9 | add 170 ml of concentrated acid (ρ = 1.84 g/ml) carefully to 750 ml of water, allow to cool and dilute to 1000 ml; commercially available (3.6.1) |
| Sodium hydroxide, c(NaOH) = 4 mol/l | 1310-73-2 | dissolve 160 g of NaOH in water and dilute to 1000 ml (3.7) |
| Manganese standard solution, 100 mg/l | — | dissolve 308 mg of manganese(II) sulfate monohydrate in water in a 1000 ml flask, add 10 ml of 3 mol/l sulfuric acid and make up to the mark; 1 ml contains 0.1 mg of Mn; commercial standard solutions may be used (3.8) |
| Water | — | deionized or distilled in an all-glass apparatus, with a manganese content as low as possible (clause 3) |
Health and safety (OHS)
- The formaldoxime solution, the 6 mol/l hydroxylammonium chloride solution and its mixture with ammonia are special hazards — hazardous operations are carried out in a fume cupboard (warning in clause 3)
- Avoid ingestion and inhalation of vapours and protect hands, eyes and face — gloves and goggles must be worn and any skin contamination washed off immediately; inhalation of formaldehyde and formaldoxime vapours causes severe irritation and oedema of the upper respiratory tract (clause 3)
- Toxic gases may be released when samples are acidified (clause 5)
Frequently asked questions
Which standard describes this test?
The test is performed according to ISO 6333:1986 — “Water quality — Determination of manganese — Formaldoxime spectrometric method”.
How does this method work?
A formaldoxime solution is added to the test portion; manganese forms an orange-red complex whose absorbance is measured at about 450 nm — at the absorption maximum (molar absorption coefficient 11 × 10³ l/(mol·cm)). The complex is stable between pH 9.
What is the measuring range and accuracy?
Range: 0.01–5 mg/l; higher concentrations after dilution (clause 1); Wavelength: about 450 nm; measured against water; Cells: 100 mm for the range 0–0.5 mg/l, 10 mm for the range 0–5 mg/l; pH range of complex stability: 9.5–10.5.
How long does the test take?
The procedure comprises 8 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Spectrometer, Glass bottles, 100 ml, Autoclave or pressure cooker, Glassware and containers, 50 ml and 1000 ml volumetric flasks, pipettes.
Where is this test used?
Water intended for human consumption — manganese 50 µg/l (Journal of Laws 2026, item 748); Surface waters (clause 1); Total manganese, including suspended and organically bound manganese — after pretreatment (6.2); Not for highly contaminated waters such as industrial waste water (clause 1).
What safety precautions apply?
The formaldoxime solution, the 6 mol/l hydroxylammonium chloride solution and its mixture with ammonia are special hazards — hazardous operations are carried out in a fume cupboard (warning in clause 3); Avoid ingestion and inhalation of vapours and protect hands, eyes and face — gloves and goggles must be worn and any skin contamination washed off immediately; inhalation of formaldehyde and formaldoxime vapours causes severe irritation and oedema of the upper respiratory tract (clause 3); Toxic gases may be released when samples are acidified (clause 5).
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number ISO 6333.