⛽ Total Base Number (TBN)
In short
Determination of the base number of petroleum products by potentiometric titration with perchloric acid in glacial acetic acid, in a solvent of toluene, acetic acid and acetone, with a pH-Ag/AgCl glass electrode for non-aqueous media. The test is performed according to PN-ISO 3771:2012 (Polish version; adopts ISO 3771:2011 [IDT]); Range (precision): unused oils 3–45, additive concentrates 5–45, used oils 3–30 mg KOH/g (clause 1). The procedure comprises 7 steps; it is used for: Unused oils — base number 3–45 (precision range, clause 1), Additive concentrates — base number 5–45 (clause 1), Used oils — base number 3–30 (clause 1); condemning limits are established empirically (ASTM D2896).
At a glance
- Standard: PN-ISO 3771:2012 (Polish version; adopts ISO 3771:2011 [IDT])
- Category: Petrochemistry
- Procedure steps: 7
- Range (precision): unused oils 3–45, additive concentrates 5–45, used oils 3–30 mg KOH/g (clause 1)
- Titrant: perchloric acid 0.1 mol/l in glacial acetic acid (5.10)
- Solvent: toluene : acetic acid : acetone = 6 : 3 : 1 (by volume) (5.9)
Overview
STATUS (as of 30.10.2026). PN-ISO 3771:2012 — Polish version, published 05.11.2012, implements ISO 3771:2011 [IDT]; it replaced PN-C-04163:1976. The PKN catalogue card has no withdrawal note — the standard is current; Polish title according to the card: “Przetwory naftowe — Oznaczanie liczby zasadowej — Metoda potencjometrycznego miareczkowania kwasem nadchlorowym”. ISO 3771:2011 (third edition) replaced ISO 3771:1994 after a technical revision.
SCOPE AND DEFINITION (clauses 1, 3). Determination of basic constituents in petroleum products — organic and inorganic bases, amino compounds, salts of weak acids (e.g. soaps), basic salts of polyacid bases and salts of heavy metals. The base number is the quantity of perchloric acid, expressed as the equivalent number of milligrams of potassium hydroxide, required to titrate 1 g of sample dissolved in the specified solvent to a well-defined inflection point. Precision was established for unused oils with base numbers 3–45, additive concentrates 5–45 and used oils 3–30.
SOLVENT AND ELECTRODE (5.9, 6.2, 8.3). Titration solvent: 6 volumes of toluene, 3 volumes of glacial acetic acid and 1 volume of acetone — the 2011 edition does not use chlorobenzene. Combination pH-Ag/AgCl electrode for non-aqueous media: low membrane resistance (100–200 MΩ), flexible ground-sleeve diaphragm, shielded reference electrode, non-aqueous electrolyte (e.g. 0.4 mol/l tetraethylammonium bromide in ethylene glycol or a saturated solution of lithium chloride in ethanol). The electrode is checked with pH 4 and pH 7 buffers — a potential difference of at least 150 mV.
STANDARDIZATION OF THE TITRANT (5.10). The 0.1 mol/l perchloric acid solution in acetic acid can be bought or prepared from concentrated perchloric acid, acetic acid and acetic anhydride (Table 1) and stored for at least 24 h. The titre is determined against tris(hydroxymethyl)aminomethane or potassium hydrogen phthalate for each new solution, at least once a week; certified solutions may be used instead of standardization.
WHAT WE DO NOT GIVE HERE. The publicly available text of the standard ends with the choice of sample mass (Table 2, 9.1). We do not give the further course of the titration, the endpoint determination, the calculation (clause 10) or the precision. The American method ASTM D2896 on the same subject states that the base number is sometimes used as a measure of lubricant degradation in service, but condemning limits must be established empirically; that method uses chlorobenzene and gives a chlorobenzene-free solvent as an alternative with poorer precision.
Method principle
The sample is dissolved in an essentially anhydrous mixture of toluene, acetone and glacial acetic acid and titrated with a standard volumetric solution of perchloric acid in glacial acetic acid using a potentiometric titrimeter with a combination pH-Ag/AgCl electrode. The meter readings are plotted against the volume of titrant, and the endpoint is taken as the last inflection of the curve (clause 4).
Applications
- Unused oils — base number 3–45 (precision range, clause 1)
- Additive concentrates — base number 5–45 (clause 1)
- Used oils — base number 3–30 (clause 1); condemning limits are established empirically (ASTM D2896)
Key parameters
| Parameter | Value |
|---|---|
| Range (precision) | unused oils 3–45, additive concentrates 5–45, used oils 3–30 mg KOH/g (clause 1) |
| Titrant | perchloric acid 0.1 mol/l in glacial acetic acid (5.10) |
| Solvent | toluene : acetic acid : acetone = 6 : 3 : 1 (by volume) (5.9) |
| Electrode | combination pH-Ag/AgCl for non-aqueous media, membrane resistance 100–200 MΩ (6.2) |
| Sample mass | 2.00 g (3–15), 1.00 g (15–30), 0.25 g (30–45) — so that about 5 ml of titrant is used at the inflection (Table 2) |
| Endpoint | last inflection of the curve (clause 4) |
Standard
- Standard number
- PN-ISO 3771:2012 (Polish version; adopts ISO 3771:2011 [IDT])
- Title (PL)
- Przetwory naftowe — Oznaczanie liczby zasadowej — Metoda potencjometrycznego miareczkowania kwasem nadchlorowym
- Title (EN)
- Petroleum products — Determination of base number — Perchloric acid potentiometric titration method
Step-by-step procedure
-
Preparing the titrant
Perchloric acid is dissolved in 500 ml of acetic acid, acetic anhydride is added in the amount from Table 1 and the solution is made up to 1 l with acetic acid; it is stored for at least 24 h (5.10.1).
-
Standardization
About 0.05 g of tris(hydroxymethyl)aminomethane or 0.1–0.2 g of potassium hydrogen phthalate, weighed to 0.1 mg, is dissolved in 60 ml of acetic acid and titrated with the same electrode system and the same endpoint method as the samples; the titre is determined for each new solution, at least once a week (5.10.2–5.10.3).
-
Preparing the electrode
The electrolyte is topped up as high as possible, and measurements are made with the filling plug removed; a new electrode is conditioned in the solvent for 12 h. If in doubt, the electrode is checked with pH 4 and 7 buffers — a difference of at least 150 mV (8.1, 8.3).
-
Preparing the sample
The sample must be representative — sediment may be acidic or basic; if necessary the sample is warmed to 60 °C to aid mixing, and used oils are shaken vigorously (clause 7).
-
Test portion
The sample mass is chosen so that about 5 ml of titrant is used at the inflection: 2.00 g for base number 3–15, 1.00 g for 15–30, 0.25 g for 30–45 (Table 2).
-
Titration
The sample dissolved in the solvent is titrated with perchloric acid while the potentiometric curve is recorded; the endpoint is the last inflection. The further procedure and the calculation are in clauses 9–10 of the standard, outside the publicly available text (clause 4).
-
Cleaning and storing the electrode
Between titrations the electrode is thoroughly washed with solvent — washing with toluene, then petroleum ether, then toluene again is effective; no oil film may remain. Before long-term storage the electrode is cleaned and kept in the same electrolyte that is used to fill it, with the filling plug closed (8.2, 8.4).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Potentiometric titrimeter | with automatic burettes accurate to at least 1/5000 of the cylinder volume; incremental or dynamic dosing — dynamic gives the best distribution of points (6.1) | — |
| Combination pH-Ag/AgCl electrode | for non-aqueous media, membrane resistance about 100–200 MΩ, flexible ground-sleeve diaphragm, internally shielded Ag/AgCl reference electrode (6.2) | — |
| Stirrer | mechanical or electrical with variable speed, with a propeller or paddle of chemically inert material; electrical stirrers earthed; a magnetic stirrer is permitted (6.3) | — |
| Titration beaker and stand | tall-form beaker; stand supporting the beaker, electrodes, stirrer and burette (6.4–6.5) | — |
| Balance | resolution 0.001 g for base numbers up to 30 and 0.0001 g above 30 (6.6) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Glacial acetic acid | 64-19-7 | component of the solvent and the titrant (5.1) |
| Acetic anhydride | 108-24-7 | for preparing the titrant (Table 1); an excess is avoided to prevent acetylation of primary or secondary amines in the sample (5.2, 5.10.1) |
| Toluene | 108-88-3 | component of the solvent (5.3) |
| Acetone | 67-64-1 | component of the solvent (5.4) |
| Perchloric acid | 7601-90-3 | 0.1 mol/l solution in acetic acid — commercial or prepared from concentrated acid (e.g. 8.5 ml of 70–72 % acid and 30 ml of acetic anhydride per 1 l); stored for at least 24 h before standardization (5.10.1) |
| Tris(hydroxymethyl)aminomethane | 77-86-1 | for standardization: dried over silica gel in a desiccator for at least 24 h; about 0.05 g in 60 ml of acetic acid (5.5, 5.10.2) |
| Potassium hydrogen phthalate | 877-24-7 | for standardization: dried for 2 h at 120 °C; 0.1–0.2 g in 60 ml of acetic acid (5.6, 5.10.2) |
| Electrode electrolyte | — | non-aqueous: tetraethylammonium bromide 0.4 mol/l in ethylene glycol or saturated lithium chloride in ethanol (70 g LiCl in 500 ml) (5.7–5.8) |
Health and safety (OHS)
- The standard warns that its use may involve hazardous materials, operations and equipment, and that the user establishes the safety and health practices (warning before clause 1)
- The 0.1 mol/l perchloric acid solution is not harmful under the test conditions, but concentrated perchloric acid is a powerful oxidant when dry or heated — contact with organic matter under such conditions is avoided and spills are washed off immediately with plenty of water (5.10.1)
- Toluene, acetone, acetic acid and acetic anhydride — according to their safety data sheets
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-ISO 3771:2012 (Polish version; adopts ISO 3771:2011 [IDT]) — “Petroleum products — Determination of base number — Perchloric acid potentiometric titration method”.
How does this method work?
The sample is dissolved in an essentially anhydrous mixture of toluene, acetone and glacial acetic acid and titrated with a standard volumetric solution of perchloric acid in glacial acetic acid using a potentiometric titrimeter with a combination pH-Ag/AgCl electrode. The meter readings are plotted against the volume of titrant, and the endpoint is taken as the last inflection of the curve (clause 4).
What is the measuring range and accuracy?
Range (precision): unused oils 3–45, additive concentrates 5–45, used oils 3–30 mg KOH/g (clause 1); Titrant: perchloric acid 0.1 mol/l in glacial acetic acid (5.10); Solvent: toluene : acetic acid : acetone = 6 : 3 : 1 (by volume) (5.9); Electrode: combination pH-Ag/AgCl for non-aqueous media, membrane resistance 100–200 MΩ (6.2).
How long does the test take?
The procedure comprises 7 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Potentiometric titrimeter, Combination pH-Ag/AgCl electrode, Stirrer, Titration beaker and stand, Balance.
Where is this test used?
Unused oils — base number 3–45 (precision range, clause 1); Additive concentrates — base number 5–45 (clause 1); Used oils — base number 3–30 (clause 1); condemning limits are established empirically (ASTM D2896).
What safety precautions apply?
The standard warns that its use may involve hazardous materials, operations and equipment, and that the user establishes the safety and health practices (warning before clause 1); The 0.1 mol/l perchloric acid solution is not harmful under the test conditions, but concentrated perchloric acid is a powerful oxidant when dry or heated — contact with organic matter under such conditions is avoided and spills are washed off immediately with plenty of water (5.10.1); Toluene, acetone, acetic acid and acetic anhydride — according to their safety data sheets.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-ISO 3771.