☠️ Cyanides in water
In short
Determination of total cyanide in water according to ISO 6703-1 — liberation of hydrogen cyanide by heating with hydrochloric acid in the presence of copper(I) ions, transfer by a current of air into sodium hydroxide solution and determination of cyanide ions by one of three methods: photometric with pyridine/barbituric acid or titrimetric. The test is performed according to ISO 6703-1:1984; Field of application: water containing less than 100 mg/l of cyanide; higher concentrations after dilution (clause 1). The procedure comprises 4 steps; it is used for: Water — total cyanide at concentrations below 100 mg/l, higher after dilution (clause 1), Municipal sewage and industrial effluents — according to the introduction to the standard, Water intended for human consumption — cyanides 50 µg/l, the method is to cover cyanide in all forms (Journal of Laws 2026, item 748).
At a glance
- Standard: ISO 6703-1:1984
- Category: Physicochemistry
- Procedure steps: 4
- Field of application: water containing less than 100 mg/l of cyanide; higher concentrations after dilution (clause 1)
- Photometric method with pyridine/barbituric acid: 0.002–0.025 mg of cyanide in the portion (clause 1)
- Titrimetric methods: using the Tyndall effect — more than 0.005 mg, using an indicator — more than 0.05 mg of cyanide (clause 1)
Overview
STATUS (as of 29.10.2026). ISO 6703-1:1984 (first edition, 01.08.1984) is published in the ISO catalogue, with no newer edition. The standard has not been adopted into the set of Polish Standards: the PKN search for "ISO 6703" returns only PN-ISO 16703:2009 and PN-EN ISO 16703:2011 (hydrocarbons — a different subject), so a reference to "PN-ISO 6703" or "PN-EN ISO 6703" points to a document that does not exist. In the ISO 6703 series, part 2 deals with easily liberatable cyanide and part 3 with cyanogen chloride (both of 1984, published), while part 4 (determination of cyanide by diffusion at pH 6, 1985) is withdrawn. Polish regulations specify other documents for cyanides (see LEGAL REQUIREMENTS) — the method described below is not interchangeable with them.
SCOPE (clause 1). Three methods for determining total cyanide in water containing less than 100 mg of cyanide per litre; higher concentrations — after dilution of the sample. Ranges of cyanide mass in the portion: photometric method with pyridine/barbituric acid — 0.002–0.025 mg, titrimetric method using the Tyndall effect — more than 0.005 mg, titrimetric method using an indicator — more than 0.05 mg. According to the introduction, the methods of parts 1–3 are suitable for controlling the quality of water and for examining municipal sewage and industrial effluents.
TOTAL CYANIDE (clause 2). Simple and complex bound cyanides, including organic compounds containing cyanogen groups that form hydrogen cyanide under the conditions of the method; cyanohydrins are detected in part. Mononitriles (R-CN), cyanate and thiocyanate ions and cyanogen chloride are not included. Complex cobalt cyanides, depending on their concentration, are decomposed only to the extent of 5–15 %, as are some organocyanide compounds (clause 3, note 1).
INTERFERENCES (clause 1, Table 1). Many ions and compounds interfere; Table 1 (not exhaustive) gives the concentrations up to which they — singly or in combination — do not interfere with the separation of hydrogen cyanide, e.g. thiosulfate, thiocyanate, carbonate and cyanate ions 1000 mg/l, nitrate and nitrite ions 500 mg/l, ammonium ions 2000 mg/l, iron(II) and iron(III) ions 5000 mg/l, silver and mercury ions 50 mg/l, chromate ions 300 mg/l, elemental chlorine 250 mg/l, hydrogen peroxide and perborate ions 10 mg/l. If any of these concentrations is likely to be exceeded, the sample is diluted with distilled water before stabilization. Aldehydes, e.g. formaldehyde, give low results (formation of cyanohydrins); fatty acids and elemental sulfur may also interfere.
LEGAL REQUIREMENTS. Water intended for human consumption: cyanides 50 µg/l (Regulation of the Minister of Health of 22 May 2026, Journal of Laws 2026, item 748, Annex 1 Part B item 16); measurement uncertainty up to 30 % of the parametric value, and the method is to determine the total amount of cyanide in all forms (Annex 4 Part B item 21 and note 3). Waste water: the reference methods for cyanides are PN-C-04603-1 (spectrophotometry), PN-C-04603-2 (titrimetry) and PN-EN ISO 14403 (continuous flow analysis) — ISO 6703-1 is not referenced there (Journal of Laws 2019, item 1311, Annex 12 item 20). Monitoring of surface and ground waters: for free and bound cyanides the regulation lists PN-80/C-04603.01, ISO 14403 and laboratory procedures (Journal of Laws 2021, item 1576, Annex 8, indicators 3.6.12–3.6.13).
WHAT WE DO NOT GIVE HERE. The publicly available text of the standard ends at the beginning of clause 6 (sampling and preservation). The rest of the preservation, the course of hydrogen cyanide liberation (clause 7) and the reagents, wavelengths, calculations and precision data of the three final determination methods (sections two to four) lie outside it, so we do not give them. For comparison: in part 3 of the same standard (cyanogen chloride) the pyridine/barbituric acid dye is measured at 578 nm.
Method principle
Liberation (section one, clause 3): the sample is heated with hydrochloric acid in the presence of copper(I) ions; the liberated hydrogen cyanide is carried by a current of air into an absorption vessel containing sodium hydroxide solution. Tin(II) chloride is added as a reducing agent, a zinc salt to provide stable zinc hexacyanoferrates, and cadmium salts as a sulfide acceptor and because of their bactericidal effect (footnote to clause 4).
Determination of cyanide ions in the absorption solution — one of three methods, because none is applicable in all cases (introduction): photometric with pyridine/barbituric acid (section two), titrimetric using the Tyndall effect (section three) or titrimetric using an indicator (section four).
Applications
- Water — total cyanide at concentrations below 100 mg/l, higher after dilution (clause 1)
- Municipal sewage and industrial effluents — according to the introduction to the standard
- Water intended for human consumption — cyanides 50 µg/l, the method is to cover cyanide in all forms (Journal of Laws 2026, item 748)
Key parameters
| Parameter | Value |
|---|---|
| Field of application | water containing less than 100 mg/l of cyanide; higher concentrations after dilution (clause 1) |
| Photometric method with pyridine/barbituric acid | 0.002–0.025 mg of cyanide in the portion (clause 1) |
| Titrimetric methods | using the Tyndall effect — more than 0.005 mg, using an indicator — more than 0.05 mg of cyanide (clause 1) |
| Liberation | heating with hydrochloric acid in the presence of copper(I), transfer of hydrogen cyanide by a current of air into NaOH solution (clause 3) |
| Not included | mononitriles, cyanate, thiocyanate, cyanogen chloride; cobalt complexes only to 5–15 % (clauses 2–3) |
| Interferences | Table 1 — limiting concentrations; aldehydes give low results (clause 1) |
Standard
- Standard number
- ISO 6703-1:1984
- Title (PL)
- Jakość wody — Oznaczanie cyjanków — Część 1: Oznaczanie cyjanków ogólnych
- Title (EN)
- Water quality — Determination of cyanide — Part 1: Determination of total cyanide
Step-by-step procedure
-
Sampling and preservation
If the sample contains undissolved cyanides, ensure a homogeneous distribution of the suspended matter in the sample and its dilutions. Immediately after sampling add to each litre of sample (or diluted sample) 5 ml of 5 mol/l sodium hydroxide, 10 ml of phenolphthalein solution and 5 ml of tin(II) chloride solution; further preservation steps are described in the part of clause 6 outside the available text.
-
Dilution in the case of interferences
If the concentration of any substance in Table 1 may exceed its limiting value, the sample is diluted with distilled water before stabilization (clause 1).
-
Liberation of hydrogen cyanide
In the apparatus of Figure 1: heating the sample with hydrochloric acid in the presence of copper(I) ions and transfer of hydrogen cyanide by a current of air into the absorption vessel with sodium hydroxide solution (clause 3); volumes, flow and time — clause 7, outside the available text.
-
Determination of cyanide ions
In the absorption solution by one of three methods, depending on the expected mass of cyanide: photometric with pyridine/barbituric acid (0.002–0.025 mg), titrimetric using the Tyndall effect (more than 0.005 mg) or titrimetric using an indicator (more than 0.05 mg) (clause 1).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Apparatus for separating hydrogen cyanide by stripping | according to Figure 1 of the standard or equivalent: 500 ml three-necked distillation flask (joints 29/32 and 14.5/23), Liebig reflux condenser, absorption vessels protected against return of liquid, funnel, flowmeter, 250 ml wash bottle for purifying the air (5.1) | — |
| pH meter | with a glass electrode fitting into the side neck of the flask (5.2) | — |
| Volumetric flasks | 25, 50, 250 and 1000 ml (5.3) | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Hydrochloric acid, ρ = 1.12 g/ml, and 1 mol/l | 7647-01-0 | for liberating hydrogen cyanide and preparing the tin(II) chloride solution (4.1–4.2) |
| Sodium hydroxide, 1 mol/l and 5 mol/l | 1310-73-2 | 4.3–4.4; the 5 mol/l solution is added to the sample immediately after sampling (clause 6) |
| Tin(II) chloride solution | — | dissolve 50 g of SnCl₂·2H₂O in 40 ml of 1 mol/l hydrochloric acid and dilute with water to 100 ml; prepare weekly (4.5) |
| Phenolphthalein solution containing chloroform | — | 0.03 g of phenolphthalein in 90 ml of ethanol, add 10 ml of chloroform (4.6) |
| Salt solutions | — | zinc and cadmium sulfate (100 g each of ZnSO₄·7H₂O and 3CdSO₄·8H₂O in 1000 ml), copper sulfate (200 g of CuSO₄·5H₂O in 1000 ml), cadmium acetate (300 g of the dihydrate in 1000 ml) (4.7–4.9) |
| Buffer solution of pH 5.4 | — | 6 g of sodium hydroxide in about 50 ml of water, add 11.8 g of succinic acid and dilute to 100 ml (4.10) |
| Water and reagents | — | distilled or deionized water, reagents of analytical grade (clause 4); reagents of the final determination methods — sections two to four, outside the available text |
Health and safety (OHS)
- Cyanides are highly toxic — carry out all operations in a fume cupboard, avoid contact with skin and eyes, always pipette with a safety pipette (pipette by bulb) (warning of the standard)
- Samples and solutions containing cyanides or heavy metals are detoxified in accordance with local official regulations (warning of the standard)
- Other reagents of the standard are also hazardous, e.g. pyridine; cadmium salt solutions and chloroform — according to the safety data sheets
Frequently asked questions
Which standard describes this test?
The test is performed according to ISO 6703-1:1984 — “Water quality — Determination of cyanide — Part 1: Determination of total cyanide”.
How does this method work?
Liberation (section one, clause 3): the sample is heated with hydrochloric acid in the presence of copper(I) ions; the liberated hydrogen cyanide is carried by a current of air into an absorption vessel containing sodium hydroxide solution. Tin(II) chloride is added as a reducing agent, a zinc salt to provide stable zinc hexacyanoferrates, and cadmium salts as a sulfide acceptor and because of their bactericidal effect (footnote to clause 4).
What is the measuring range and accuracy?
Field of application: water containing less than 100 mg/l of cyanide; higher concentrations after dilution (clause 1); Photometric method with pyridine/barbituric acid: 0.002–0.025 mg of cyanide in the portion (clause 1); Titrimetric methods: using the Tyndall effect — more than 0.005 mg, using an indicator — more than 0.05 mg of cyanide (clause 1); Liberation: heating with hydrochloric acid in the presence of copper(I), transfer of hydrogen cyanide by a current of air into NaOH solution (clause 3).
How long does the test take?
The procedure comprises 4 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Apparatus for separating hydrogen cyanide by stripping, pH meter, Volumetric flasks.
Where is this test used?
Water — total cyanide at concentrations below 100 mg/l, higher after dilution (clause 1); Municipal sewage and industrial effluents — according to the introduction to the standard; Water intended for human consumption — cyanides 50 µg/l, the method is to cover cyanide in all forms (Journal of Laws 2026, item 748).
What safety precautions apply?
Cyanides are highly toxic — carry out all operations in a fume cupboard, avoid contact with skin and eyes, always pipette with a safety pipette (pipette by bulb) (warning of the standard); Samples and solutions containing cyanides or heavy metals are detoxified in accordance with local official regulations (warning of the standard); Other reagents of the standard are also hazardous, e.g. pyridine; cadmium salt solutions and chloroform — according to the safety data sheets.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number ISO 6703-1.