⚗️ Bromate in Water — Ion Chromatography

Physicochemistry PN-EN ISO 15061

In short

Determination of dissolved bromate (BrO₃⁻) in water by ion chromatography with conductivity detection after chemical suppression according to PN-EN ISO 15061. The test is performed according to PN-EN ISO 15061:2003; Range of applicability: 0.5–1000 µg/l of bromate after appropriate sample pretreatment (clause 1, 4.2). The procedure comprises 8 steps; it is used for: Water intended for human consumption — parametric value for bromate 10 µg/l (Dz.U. 2026 item 748, Annex 1 part B item 11), Raw, surface and partially treated water (clause 1), Swimming pool water (clause 1).

At a glance

  • Standard: PN-EN ISO 15061:2003
  • Category: Physicochemistry
  • Procedure steps: 8
  • Range of applicability: 0.5–1000 µg/l of bromate after appropriate sample pretreatment (clause 1, 4.2)
  • Detection: conductivity with chemical suppression; UV 190–205 nm only for confirmation (4.4, clause 5)
  • Resolution: R ≥ 1.3 between bromate and the nearest peak, usually chloride (clauses 5 and 8)

Overview

STATUS (as of 03.11.2026). PN-EN ISO 15061:2003 — Polish version, published on 29.08.2003, implements EN ISO 15061:2001 and ISO 15061:2001 [IDT]; it replaced the English version PN-EN ISO 15061:2002. The PKN catalogue card has no note of withdrawal. ISO 15061:2001 (first edition) is published in the ISO catalogue with the stage “to be revised” (90.92); the iTeh catalogue shows no newer edition or draft. The standard covers bromate only — bromide is determined according to PN-EN ISO 10304-1 (a separate method in the Knowledge Base: dissolved anions in water), and the method with post-column reaction is described in PN-EN ISO 11206.

SCOPE (clause 1). Dissolved bromate in water, e.g. drinking water, raw water, surface water, partially treated water and swimming pool water. Appropriate pretreatment of the sample — elimination of chloride, sulfate and metals, preconcentration or dilution — gives a range of applicability from 0.5 µg/l to 1000 µg/l of bromate. The working range is restricted by the ion-exchange capacity of the concentrator column and of the separator column; if necessary, the sample is diluted.

PRINCIPLE (clause 4). Sample pretreatment removes ozone and solids and reduces chloride, sulfate, carbonate, hydrogen carbonate and metals by means of cation exchangers. Bromate is separated by liquid chromatography on an anion exchanger — on the separator column or after elution from a concentrator column; the eluents are usually aqueous solutions of salts of weak mono- and dibasic acids. It is measured with a conductivity detector with chemical suppression, while a UV detector (190–205 nm) is suitable only to confirm the result. Strongly retained ions (nitrate, phosphate, sulfate) are removed from the column with a more concentrated eluent; the concentration is determined after calibration of the overall procedure.

ESSENTIAL MINIMUM REQUIREMENTS (clauses 5 and 8). Recovery of bromate with preconcentration 80–120 %. Peak resolution between bromate and the nearest peak (usually chloride) of at least 1.3, calculated as R = 2(tR2 − tR1)/(w1 + w2), where tR are the retention times and w the peak widths at the base. Conductivity detection with chemical suppression, calibration according to ISO 8466-1 or ISO 8466-2, a validity check of the calibration function; if necessary, replicate determinations and, with matrix interferences, the method of standard addition.

INTERFERENCES (clause 3). Nitrate, chloride, carbonate and sulfate may reduce the capacity of the concentrator column and the recovery of bromate; chloride, sulfate, carbonate and hydrogen carbonate interfere with the determination, and depending on the column also other ions. Metals — e.g. barium and silver ions released during sample pretreatment — bind to the column packing; they are eliminated with a metal clean-up column or special exchangers. Solids and organic compounds (mineral oils, detergents, humic acids) shorten the lifetime of the columns; the effect of the organic acids checked was found not to be significant (Annex E).

REAGENTS AND SOLUTIONS (clause 6). Water of grade 1 according to ISO 3696. Bromate stock solution 1000 mg/l BrO₃⁻: 1.306 g of potassium bromate dried for at least 1 h at (105 ± 5) °C is dissolved in a 1000 ml volumetric flask, stored at 2–6 °C and renewed every 12 months (alternatively a commercial solution). Standard solutions: I — 100 mg/l (renewed every 6 months), II — 1 mg/l (every 3 months). There are 5–10 calibration solutions, prepared on the day of use, e.g. for the range 0.5–5.0 µg/l from solution II. Two eluents are used: a separation eluent (lower concentration) and a flushing eluent (higher — removes strongly retained ions); the eluents are degassed, stored in the dark and renewed every 3 days, and their composition is chosen for the column and detector.

FORMATION AND REGULATIONS. Bromate may form during ozonation of water containing bromide — in a direct reaction of ozone with bromide or via radicals; IARC has classified it in group 2B as possibly carcinogenic (textbook “Chemia wody” [Water chemistry], Wrocław University of Science and Technology 2024, section 5.6). The parametric value for bromate in water intended for human consumption is 10 µg/l (Regulation of the Minister of Health of 22 May 2026, Dz.U. 2026 item 748, Annex 1 part B item 11); the method must have a limit of quantification of no more than 30 % of the parametric value and a measurement uncertainty of no more than 40 % (Annex 4 part B). The regulation applies from 24.06.2026; the previous regulation of 7.12.2017 (Dz.U. 2017 item 2294) ceased to have effect on that day.

WHAT WE DO NOT GIVE HERE. The publicly available text of the standard ends with clause 8 (quality requirements for the separator column). Sampling and sample pretreatment (clause 9 — removal of ozone, cartridges), the procedure with preconcentration (clause 10), the calculation of the result, examples of eluents and regeneration solutions (Annexes A, B), an example configuration with preconcentration (Annex C) and the interlaboratory trial (Annex D) are outside it — which is why we do not give the composition of eluents, sample volumes, retention times, limits of detection or precision data.

Method principle

Bromate is separated from the other anions on an anion exchanger by ion chromatography and measured with a conductivity detector with chemical suppression. Before the measurement ozone and solids are removed from the sample, and the content of chloride, sulfate, carbonate and metals is reduced on cartridges with cation exchangers. At low concentrations bromate is preconcentrated on a concentrator column, and the concentration is determined from the calibration function of the overall procedure (ISO 8466-1 or ISO 8466-2); the UV detector serves only to confirm the result.

Applications

Key parameters

ParameterValue
Range of applicability0.5–1000 µg/l of bromate after appropriate sample pretreatment (clause 1, 4.2)
Detectionconductivity with chemical suppression; UV 190–205 nm only for confirmation (4.4, clause 5)
ResolutionR ≥ 1.3 between bromate and the nearest peak, usually chloride (clauses 5 and 8)
Recovery with preconcentration80–120 % (clause 5)
Calibration5–10 solutions prepared on the day of use; linear (ISO 8466-1) or second-order (ISO 8466-2) function (clause 5, 6.13)
Parametric value in drinking water10 µg/l; limit of quantification ≤ 30 %, uncertainty ≤ 40 % of the parametric value (Dz.U. 2026 item 748, Annex 1 part B item 11, Annex 4 part B)

Standard

Standard number
PN-EN ISO 15061:2003
Title (PL)
Jakość wody — Oznaczanie bromianów rozpuszczonych — Metoda chromatografii jonowej
Title (EN)
Water quality — Determination of dissolved bromate — Method by liquid chromatography of ions

Step-by-step procedure

  1. Sampling

    Take the sample according to ISO 5667-1, ISO 5667-2 and ISO 5667-3 (normative references); preservation and storage are described in clause 9, which is outside the available text of the standard.

  2. Sample pretreatment

    Remove ozone and solids, and reduce chloride, sulfate, carbonate, hydrogen carbonate and metals on cartridges with cation exchangers; remove metals released from the cartridges with a metal clean-up column (3.3, 4.1, 7.2).

  3. Eluents

    Prepare a separation eluent (lower concentration) and a flushing eluent (higher concentration) suitable for the column and detector; degas them, avoid renewed air pick-up (e.g. by helium sparging), store them in the dark and renew every 3 days (6.10).

  4. Calibration

    On the day of use prepare 5–10 calibration solutions distributed evenly over the working range — e.g. 0.5–5.0 µg/l from standard solution II — and establish the calibration function of the overall procedure according to ISO 8466-1 or ISO 8466-2 (4.6, 6.13).

  5. Resolution check

    In the chromatograms of samples and standard solutions check that the resolution between the bromate peak and the nearest peak (usually chloride) is at least 1.3: R = 2(tR2 − tR1)/(w1 + w2) (clause 8).

  6. Measurement

    Introduce the sample onto the separator column directly or, at low concentrations, via the concentrator column (recovery 80–120 %); measure with the conductivity detector with suppression, and if necessary confirm the result with the UV detector (4.2–4.4, clause 5).

  7. Column flushing

    Remove strongly retained ions (e.g. nitrate, phosphate, sulfate) from the column with the higher-concentration flushing eluent (4.5, 6.10.2).

  8. Quality control

    Check the validity of the calibration function; if necessary carry out replicate determinations, and with matrix interferences use the method of standard addition (clause 5).

Required equipment and apparatus

EquipmentExampleIndicative price
Ion chromatographeluent reservoirs and degassing for two eluents, pump suitable for the step gradient technique, injection system with a loop of appropriate volume (e.g. 0.05–2 ml) or an autosampler, column-switching valves with control (7.1 a–d)—
Separator column and concentrator columnseparator column meeting the resolution requirement of clause 8; concentrator column — for low concentrations (7.1 e–f)—
Detectorsconductivity detector with an anion suppressor device; UV (e.g. spectrophotometer 190–400 nm) to confirm the result (7.1 g–h)—
Cartridges for sample pretreatmentcation exchangers in the Ag, Ba and H form; optionally an on-line metal clean-up column; cartridges with non-polar phases, e.g. polyvinylpyrrolidone (7.2)—
Recording and evaluationrecorder, integrator or PC with software for data acquisition and evaluation (7.1 i)—

Reagents, media and consumables

ReagentCASDetails
Water—grade 1 according to ISO 3696; degassed for eluents (6.1, 6.10)
Potassium bromate7758-01-2stock solution 1000 mg/l BrO₃⁻: 1.306 g after drying for at least 1 h at (105 ± 5) °C per 1000 ml; storage at 2–6 °C, renewed every 12 months; alternatively a commercial solution (6.6, 6.11)
Bromate standard solutions—I — 100 mg/l (10.0 ml of stock solution per 100 ml, renewed every 6 months), II — 1 mg/l (1.0 ml of solution I per 100 ml, every 3 months); calibration solutions are prepared on the day of use (6.12, 6.13)
Sodium carbonate and sodium hydrogen carbonate497-19-8reagents of common eluents (sodium hydrogen carbonate: CAS 144-55-8); the eluent composition is chosen for the column and detector (6.2, 6.3, 6.10)
Disodium tetraborate decahydrate and boric acid1303-96-4reagents of common eluents (boric acid: CAS 10043-35-3) (6.4, 6.5)
Nitric acid, sulfuric acid, ethylenediamine—nitric acid 0.1 mol/l, sulfuric acid with a density of 1.84 g/ml and ethylenediamine — reagents listed in the standard (6.7–6.9); composition of regeneration solutions according to the column manufacturer (6.14)

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 15061:2003 — “Water quality — Determination of dissolved bromate — Method by liquid chromatography of ions”.

How does this method work?

Bromate is separated from the other anions on an anion exchanger by ion chromatography and measured with a conductivity detector with chemical suppression. Before the measurement ozone and solids are removed from the sample, and the content of chloride, sulfate, carbonate and metals is reduced on cartridges with cation exchangers.

What is the measuring range and accuracy?

Range of applicability: 0.5–1000 µg/l of bromate after appropriate sample pretreatment (clause 1, 4.2); Detection: conductivity with chemical suppression; UV 190–205 nm only for confirmation (4.4, clause 5); Resolution: R ≥ 1.3 between bromate and the nearest peak, usually chloride (clauses 5 and 8); Recovery with preconcentration: 80–120 % (clause 5).

How long does the test take?

The procedure comprises 8 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Ion chromatograph, Separator column and concentrator column, Detectors, Cartridges for sample pretreatment, Recording and evaluation.

Where is this test used?

Water intended for human consumption — parametric value for bromate 10 µg/l (Dz.U. 2026 item 748, Annex 1 part B item 11); Raw, surface and partially treated water (clause 1); Swimming pool water (clause 1); Ozonated water — bromate forms during ozonation of water containing bromide (textbook “Chemia wody”, Wrocław University of Science and Technology 2024, section 5.6).

What safety precautions apply?

Potassium bromate — harmonised CLP classification: H271 (may cause fire or explosion; strong oxidiser), H301 (toxic if swallowed), H350 (may cause cancer); The acids and ethylenediamine listed in clause 6 of the standard are handled according to their safety data sheets; Sodium carbonate — harmonised CLP classification: H319 (causes serious eye irritation).

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN ISO 15061.

🔍 Find a laboratory performing this test